Subject:
MAO/SMPS/S1 - clogged impactor at the SMPS aerosol sample inlet
DataStreams:maoaossmpsS1.a1
Description:
The clogging of the impactor leads to a decrease in the aerosol sample flow rate of the 
SMPS, which is nominally 1 lpm.  This decrease in the SMPS sample flow rate results in an 
underestimation of the ambient size distribution and integrated number concentration.
Suggestions: 
null
Measurements:maoaossmpsS1.a1:
  • Total scan concentration over all bins(total_concentration)
  • CPC inlet flow rate(CPC_inlet_flow)
  • Number size distribution (dN/dlogDp)(number_size_distribution)
  • CPC sample flow rate(CPC_sample_flow)
  • Aerosol flow rate(aerosol_flow)