DQRID : D140724.3
Start DateStart TimeEnd DateEnd Time
06/04/2014140007/24/20141934
more
Subject:
ENA/MWRP/C1 - Reprocess: Bias in brightness temperatures
DataStreams:enamwrpC1.b1
Description:
The brightness temperatures between 22.234 and 26.234 GHz show a bias of 1.7 to 3.6 
degrees K. Data needs to be recalibrated. As a consequence the retrieved precipitable water 
vapor has a positive bias.
Suggestions: 
Do not use data
Measurements:enamwrpC1.b1:
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Microwave brightness temperature(brightnessTemperature)