DQRID : D141010.6
Start DateStart TimeEnd DateEnd Time
10/05/2014190410/07/20141311
more
Subject:
MAO/TRACEGAS/S1 - CO instrument operated incorrectly
DataStreams:maoaoscoS1.a1
Description:
New operator mistakenly installed impermeable blue filter separator instead of the 
permeable white filter.
Suggestions: 
Do not use data
Measurements:maoaoscoS1.a1:
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • Measured water content(h2o)
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • Carbon monoxide concentration(co)
  • Nitrous oxide (N2O) mixing ratio calculated with nominal sensitivity correction(n2o)
more