DQRID : D141013.5
Start DateStart TimeEnd DateEnd Time
07/22/2014175010/02/20141829
more
Subject:
OLI/IRT/M1 - Instrument Communication
DataStreams:oliirtM1.b1, oliirt200msM1.a1
Description:
The serial cable connector became disconnected, apparently causing the datalogger to fail. 
The datalogger was replaced on 7/23 and the IRT analog cable was reconnected but the 
serial cable was not connected to avoid the same event happening again. The DB9 connectors 
inside the IRT enclosure were cleaned of corrosion and the serial cable was eventually 
reconnected.
Suggestions: 
Measurements:oliirtM1.b1:
  • Sky infrared temperature, standard deviation(sky_ir_temp_std)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • -180.0 - +180.0(alt)
  • lon(lon)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • Sky Infra-Red Temperature Minima(sky_ir_temp_min)
  • Internal Reference Infra Red Temperature(ref_ir_temp)
  • base time in epoch(base_time)
  • Number of averaged samples(samples_in_average)
  • time_in_seconds_since_volume_start(time)
  • Time offset from base_time(time_offset)
  • North Latitude(lat)
more
oliirt200msM1.a1:
  • Time offset from base_time(time_offset)
  • lon(lon)
  • -180.0 - +180.0(alt)
  • time_in_seconds_since_volume_start(time)
  • Internal Reference Infra Red Temperature(ref_ir_temp)
  • Sky Infra-Red Temperature(sky_ir_temp)
  • base time in epoch(base_time)
  • North Latitude(lat)
more

Close this window