DQRID : D141022.2
Start DateStart TimeEnd DateEnd Time
07/12/2012000006/19/20142000
more
Subject:
MAG/PSAP/M1 - Impactor state backwards
DataStreams:magaospsap3wM1.b1
Description:
Impactor was installed on 2012-07-12 but did not toggle until 2013-05-16 00:04:19 UTC.

The impactor as delivered did not have limit switches indicating the impactor state. Only 
the command signal was available and is saved in the 'master' data stream. A programming 
error led to the command signal having reverse polarity (1 is actually 10 and 10 is 
actually 1-um cutoff)
Suggestions: 
Alternative methods for determining the impactor state:

1) Monitor the PSAP flow.  The flow decreases slightly in the 1 um state

null
Measurements:magaospsap3wM1.b1:
  • Aerosol light absorption coefficient, nominal red wavelength at dry or reference
    RH(Ba_R_PSAP3W)
  • AOS Ba_B_CLAP3W transmittance, blue channel(transmittance_B)
  • AOS Ba_R_CLAP3W transmittance, red channel(transmittance_R)
  • Mass flow(mass_flow)
  • Diffusion flow valve position(dilution_flow_valve_position)
  • AOS Ba_B_CLAP3W_1 (NOAA) sample length, needed to calculate Beers Law(sample_length)
  • Aerosol light absorption coefficient, nominal green wavelength at dry or
    reference RH(Ba_G_PSAP3W)
  • Impactor setting in terms of aerodynamic diameter cut off(impactor_setting)
  • AOS PSAP3W transmittance, green channel(transmittance_G)
  • Flow setpoint(dilution_flow_setpoint)
  • Dilution flow rate(dilution_flow)
  • Aerosol light absorption coefficient, nominal blue wavelength at dry or
    reference RH(Ba_B_PSAP3W)