Subject: | ENA/MAG/PVC/TMP/CCN/C1 - Reprocess: CCN temperature unstable flag reversed |
DataStreams: | tmpaosccn100S1.a1, enaaosccn100C1.a1, pvcaosccn100S1.a1, pvcaosccn100M1.a1, magaosccn100M1.a1
|
Description: | The variable CCN-temp unstable is incorrectly defined as a hexadecimal value. The values
are 0 or 1 only. The description is also reversed. A value of 1 indicates the instrument
is operating stably. |
Suggestions: | Reverse the values. Change 0 to 1, and 1 to 0 |
Measurements: | tmpaosccn100S1.a1: - AOS CCN logical value, indicating whether the instrument has reached thermal
gradient stabilization and is within 0.4 degrees C of the set point(CCN_temp_unstable_flag) more
pvcaosccn100M1.a1: - AOS CCN logical value, indicating whether the instrument has reached thermal
gradient stabilization and is within 0.4 degrees C of the set point(CCN_temp_unstable_flag) more
magaosccn100M1.a1: - AOS CCN logical value, indicating whether the instrument has reached thermal
gradient stabilization and is within 0.4 degrees C of the set point(CCN_temp_unstable_flag) more
pvcaosccn100S1.a1: - AOS CCN logical value, indicating whether the instrument has reached thermal
gradient stabilization and is within 0.4 degrees C of the set point(CCN_temp_unstable_flag) more
enaaosccn100C1.a1: - AOS CCN logical value, indicating whether the instrument has reached thermal
gradient stabilization and is within 0.4 degrees C of the set point(CCN_temp_unstable_flag) more
|