DQRID : D141212.4
Start DateStart TimeEnd DateEnd Time
11/01/2014000012/12/20141815
more
Subject:
NSA/MWRP/C1 - Reprocessing Notice (RID 1262)
DataStreams:nsamwrpC1.b1
Description:
See below for details of reprocessing that was completed on 20150309.  Original DQR 
category: 4 (Red, Incorrect).

-----------------------------------------------------------

The instrument appears to be unstable and the brightness temperatures are too warm. This 
affects the retrievals as well. The temperature dependence of the calibration coefficients 
has slightly changed.
Suggestions: 
Measurements:nsamwrpC1.b1:
  • Microwave brightness temperature(brightnessTemperature)
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
  • Retrieved liquid water path(liquidWaterPath)
  • Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)