Subject: | ENA/MWRP/C1 - Reprocess: Calibration drift |
DataStreams: | enamwrpC1.b1
|
Description: | The data are incorrect mostly because of fast drift in the calibration. Brightness
temperatures become negative after 9/7/2015. |
Suggestions: | Data will be recalibrated. Wait for data to be recalibrated before using. |
Measurements: | enamwrpC1.b1: - Retrieved liquid water path(liquidWaterPath)
- Retrieved water vapor density(waterVaporDensity)
- Derived relative humidity(relativeHumidity)
- Retrieved liquid water path using only 23.835 and 30.0 GHz(liquidWaterPath2)
- Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
- Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
- Interpolated water vapor mixing ratio(waterVaporMixingRatio)
- Microwave brightness temperature(brightnessTemperature)