Subject:
SGP/CCN/C1 - Instrument under counting at lowest super saturation value
DataStreams:sgpaosccn100C1.a1, sgpaosccnavgC1.c1, sgpaosccnavgC1.c2, sgpaosccnC1.a1, sgpaosccnC1.00
Description:
The slow temperature decay of the OPC induced droplet evaporation and caused an under 
counting of the droplets at the lowest %ss, 0.15.
Suggestions: 
Use next highest %ss value, 0.25