DQRID : D150327.13
Start DateStart TimeEnd DateEnd Time
02/14/2015165802/25/20151800
more
Subject:
MAO/TRACEGAS/S1 - Data spikes
DataStreams:maoaoscoS1.a1
Description:
There are multiple dropouts of 1-3 s where the instrument appears to lose laser lock.  
These result in spikes of -1 and > 2 ppmv on the output and should be ignored.  Total amount 
of data loss is very minimal (~120 s over the two weeks)
Suggestions: 
Remove spikes before using data
Measurements:maoaoscoS1.a1:
  • Nitrous oxide (N2O) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(n2o_dry)
  • Measured water content(h2o)
  • Carbon monoxide (CO) mixing ratio corrected for water vapor concentration and
    calculated with nominal sensitivity correction(co_dry)
  • Carbon monoxide concentration(co)
  • Nitrous oxide (N2O) mixing ratio calculated with nominal sensitivity correction(n2o)
more