Subject: | SGP/RL/C1 - Alignment malfunction resulting in poor signal-to-noise ratio |
DataStreams: | sgprlC1.a0, sgprlC1.00
|
Description: | The signal-to-noise ratio was poor during the effected period due to a malfunction of the
alignment system. This was probably caused by a problem in the stepper motor that
controls the beam steering mirror. |
Suggestions: | Users should avoid using data during this period. |
Measurements: | sgprlC1.00: - (Raw data stream - documentation not supported)
more
sgprlC1.a0: - Summed analog signal in the temperature 2 channel(t2_analog_high)
- Number of photons counted in the low water vapor channel(water_counts_low)
- Summed analog signal in the low water vapor channel(water_analog_low)
- Summed analog signal in the high depolarization channel(depolarization_analog_high)
- Number of photons counted in the high nitrogen channel(nitrogen_counts_high)
- Summed analog signal in the liquid water channel(liquid_analog_high)
- Number of photons counted in the temperature 2 channel(t2_counts_high)
- Count rate in the high depolarization channel(depolarization_counts_high)
- Number of photons counted in the high elastic (aerosol) channel(elastic_counts_high)
- Count rate in the high water vapor channel(water_counts_high)
- Number of photons counted in the temperature 1 channel(t1_counts_high)
- Summed analog signal in the low nitrogen channel(nitrogen_analog_low)
- Summed analog signal in the high water vapor channel(water_analog_high)
- Summed analog signal in the high elastic channel(elastic_analog_high)
- Summed analog signal in the high nitrogen channel(nitrogen_analog_high)
- Number of photons counted in the liquid water channel(liquid_counts_high)
- Number of photons counted in the low elastic (aerosol) channel(elastic_counts_low)
- Summed analog signal in the low elastic channel(elastic_analog_low)
- Summed analog signal in the temperature 1 channel(t1_analog_high)
- Number of photons counted in the low nitrogen channel(nitrogen_counts_low)
more
|