DQRID : D150427.1
Start DateStart TimeEnd DateEnd Time
10/09/2014000012/31/20152359
more
Subject:
MAO/MWRP/M1 - Reprocess: High precipitable water vapor
DataStreams:maomwrpM1.b1
Description:
The precipitable water vapor appears to have a high bias that appears to be due to 
brightness temperatures being calibrated 1.5-2 K too warm. 
The reason for the calibration bias is not known and several attempts at correcting it 
have not been successful. The liquid water path is not affected as much.
Suggestions: 
Measurements:maomwrpM1.b1:
  • Total precipitable water vapor, from microwave radiometer(totalPrecipitableWater)
  • Retrieved total precipitable water vapor using only 23.835 and 30.0 GHz(totalPrecipitableWater2)
more