Subject:
ENA/ACSM/C1 - Bad ACSM data
DataStreams:enaaosacsmC1.a1
Description:
After initial startup of the container at ENA, the ACSM was behaving erratically.  It was 
determined that the filter switching valve had failed.  Because of the difficulties in 
getting parts to the site, the replacement valve was not delivered until June, 2014.  After 
delivery, there were tuning issues with the instrument.  Instrument tuning was 
accomplished remotely on 2014-07-10 and resulted in good data.
Suggestions: 
Do not use data
Measurements:enaaosacsmC1.a1:
  • Mass concentration of sulfate, ambient aerosol in air.(sulfate)
  • Mass concentration of nitrate, ambient aerosol in air(nitrate)
  • Mass concentration of total organics, ambient aerosol in air(total_organics)
  • Mass concentration of ammonium, ambient aerosol in air(ammonium)
  • Mass concentration of chloride, ambient aerosol in air.(chloride)