DQRID : D150513.3
Start DateStart TimeEnd DateEnd Time
02/19/2015092502/19/20151105
02/16/2015105202/17/20150524
02/17/2015055702/17/20150732
04/27/2015094204/27/20151523
05/03/2015052205/03/20150811
05/03/2015022705/03/20150522
02/20/2015191402/22/20152131
02/19/2015154102/20/20150148
05/08/2015094205/08/20151340
02/17/2015085802/17/20151135
02/18/2015095402/18/20151144
02/19/2015080802/19/20150914
04/26/2015090704/26/20151352
02/20/2015125502/20/20151829
more
Subject:
ENA/AOSMET/C1 - Instrument
DataStreams:enaaosmetC1.a1
Description:
Instrument experienced periodic dropouts.  These were apparent on all channels.  The RH 
sensor went off scale.  During these periods there are no valid points for Air Temperature, 
RH or BP. RH should be considered uncalibrated until this date.  Changing the sensor at 
the end of this period fixed the problem.
Suggestions: 
Measurements:enaaosmetC1.a1:
  • Ambient air temperature(T_Ambient)
  • Wind Direction, relative to true North(WindDirection)
  • Wind speed(WindSpeed)
  • Rain amount(rain_amount)
  • Ambient air relative humidity(RH_Ambient)
  • Rain intensity(rain_intensity)
  • Rain duration(rain_duration)
  • Pressure of ambient air(P_Ambient)