ENA/CPC/C1 - Data reprocessed to account for vacuum port too low
DataStreams:enaaoscpcfC1.a1
Description:
Recorded calibrations relating the effect on flow of reduced nozzle and orifice pressure 
was used to correct the data.

When the CPC was put back in operation, the external vacuum port was connected to a vacuum 
line, which didn't provide enough pull.