Start Date | Start Time | End Date | End Time |
---|---|---|---|
09/14/2014 | 0500 | 09/15/2015 | 1552 |
Subject: | MAO/CPC/S1 - Clogged nozzle |
DataStreams: | maoaoscpcfS1.a1 |
Description: | There is evidence that the nozzle of the CPC developed a clog that worsened gradually over time. The increased nozzle pressure reduced the flow rate, which will affect the calculation of the particle number concentration. |
Suggestions: | null |
Measurements: | maoaoscpcfS1.a1: more
|