DQRID : D150930.2
Start DateStart TimeEnd DateEnd Time
09/21/2015000009/25/20151625
more
Subject:
MAO/UHSAS/S1 - Instrument sample flow low
DataStreams:maoaosuhsasS1.a1
Description:
Instrument sample flow steadily decreasing from the nominal 50 sccm to about 13-14 sccm 
over several days due to an internal obstruction. Possible particle losses, potentially 
affecting the measured total number concentration and sizing.
Suggestions: 
Use data with caution
Measurements:maoaosuhsasS1.a1:
  • Particle size distribution(size_distribution)
  • Sample pressure(sample_pressure)
  • Main flow rate(flow)
more