DQRID : D151111.5
Start DateStart TimeEnd DateEnd Time
06/19/2015020007/07/20151600
more
Subject:
MAO/AOS/S1 - Blocked 1 micron impactor state
DataStreams:maoaosimpactorS1.a1
Description:
A bug or clump of dirt stuck on the inlet of the 1 micron impactor. This increased the air 
flow restriction in the 1 micron impactor, changing the cut off point.
Suggestions: 
The 10 micron data is correct. The 1 micron is incorrect. Do not use 1 micron data.
Measurements:maoaosimpactorS1.a1:
  • Impactor state in terms of aerodynamic diameter cut off(impactor_state)
more