DQRID : D151116.1
Start DateStart TimeEnd DateEnd Time
09/17/2015000001/25/20161436
more
Subject:
ENA/OZONE/C1 - Reprocess: Under reporting ozone concentration
DataStreams:enaaosozoneC1.a1, enaaosozoneC1.00
Description:
Switching valve developed an internal leak that gradually worsened. Reported instrument 
values are under-representations of the ambient ozone concentration during this period. The 
under reporting worsens from 85% to 77% over the course of this DQR. No corrections have 
been applied at this point.
Suggestions: 
null
Measurements:enaaosozoneC1.a1:
  • Ozone concentration(ozone)
more
enaaosozoneC1.00:
  • (Raw data stream - documentation not supported)
more

Close this window