DQRID : D151203.8
Start DateStart TimeEnd DateEnd Time
10/19/2015033210/19/20150854
10/11/2015022810/11/20151354
09/23/2015054009/23/20151055
10/07/2015045910/07/20151141
10/06/2015021710/06/20151339
10/03/2015030910/03/20151200
10/01/2015003210/01/20151947
10/18/2015052810/18/20151127
10/01/2015232010/01/20152359
10/17/2015021310/17/20150427
10/05/2015000010/05/20151525
10/17/2015111210/17/20151304
10/25/2015041510/25/20151142
10/12/2015073510/12/20151200
10/24/2015031410/24/20151130
10/04/2015000010/04/20151230
10/04/2015190110/04/20152359
09/21/2015065109/21/20151146
10/02/2015000010/02/20151530
09/12/2015081309/12/20151201
10/08/2015014310/08/20151532
10/18/2015130110/18/20151700
09/22/2015004509/22/20150801
more
Subject:
MAO/UHSAS/S1 - Incorrect concentration
DataStreams:maoaosuhsasS1.a1
Description:
Ambient aerosol concentration was above UHSAS upper detection limit, saturating the 
instrument. Instrument started showing incorrect (low) concentration values.
Suggestions: 
Do not use the data from these time periods.
Measurements:maoaosuhsasS1.a1:
  • Particle size distribution(size_distribution)
more

Close this window