Subject:
ENA/UHSAS/C1 - High counts on three lowest channels
DataStreams:enaaosuhsasC1.a1
Description:
High counts in the UHSAS three lowest channels. Probably a contamination (a fiber on the 
jet nozzle) inside the optical chamber. Rest of the channels should not be affected.
Suggestions: 
Do not use the data from the three lowest channels.