DQRID : D160308.3
Start DateStart TimeEnd DateEnd Time
11/15/2015204911/16/20152059
more
Subject:
AWR/CPC/M1 - Instrument Startup, No Data Collected
DataStreams:awraoscpcfM1.a1
Description:
At the instrument startup of the AMF2 deployment, the computers were turned on before the 
instruments were turned on.  The data collection starts automatically on instrument 
startup, but if the instruments are not on and connected, the data text files will contain 
missing data codes such as NaN or -9999.
Suggestions: 
No data available during this time period
Measurements:awraoscpcfM1.a1:
  • Aerosol concentration averaged over times to match N_CCN(concentration)
  • Condensation temperature(condensation_temp)
  • Liquid level(liquid_level)
  • Cabinet temperature(cabinet_temp)
  • Orifice pressure(orifice_pressure)
  • Saturation temperature(saturation_temp)
  • Ambient pressure(ambient_pressure)
  • OPC laser current(laser_current)
  • Nozzle pressure(nozzle_pressure)
  • Liquid fraction available(fraction_full)
  • Optics temperature(optics_temp)
more