Subject: | AWR/GNDRAD/S1 - IRT failure |
DataStreams: | awrgndrad20sS1.a0, awrgndrad60sS1.b1
|
Description: | A heater malfunctioned and overheated the IRT causing it to fail.
[The datalogger was using the wrong multiplier and offset. The values used were 26 K/V and
173.2 K, respectively (same as SKYRAD), but the correct values are 25 K/V and 223.2 K,
respectively.] |
Suggestions: | The surface temperature (sfc_ir_temp) measurements cannot be corrected. |
Measurements: | awrgndrad20sS1.a0: - Instantaneous Surface Infra-Red Signal(inst_sfc_ir_signal)
more
awrgndrad60sS1.b1: - Surface Infra-Red Temperature, Maxima(sfc_ir_temp_max)
- Surface Infra-Red Temperature, Minima(sfc_ir_temp_min)
- Surface infra-red temperature(sfc_ir_temp)
- Surface Infra-Red Temperature, Standard Deviation(sfc_ir_temp_std)
more
|