Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2000 | 0000 | 04/01/2016 | 0000 |
Subject: | SGP/NEPHELOMETER/C1 - Bad NEPH T/RH values from thermal gradients |
DataStreams: | sgpaosnephwet1mC1.b1, sgpaosnephwetC1.a1, sgpaosnephwetC1.b1 |
Description: | Due to the thermal inertia of the instrument wall, slow signal response and a radial temperature gradient inside the instrument cavity, the reported RH and temperature inside the wet nephelometer are off by 2-10% RH and 1-2 degrees C. The radial temperature gradients inside the instrument need to be characterized at several locations within the instrument with respect to a calibrated Vaisala T and RH sensor at the instrument exit. The resulting fRH VAP needs to be recalculated and analyzed. |
Suggestions: | Do not use data |
Measurements: | sgpaosnephwet1mC1.b1: more
sgpaosnephwetC1.b1: more sgpaosnephwetC1.a1: more |