Subject: | SGP/CCN/S01 - Concentration measurement differences |
DataStreams: | sgpaosccn200S01.00, sgpaosccn200S01.a1
|
Description: | Total particle number concentrations different approximately by 20-30% between the two
columns at the same supersaturation setpoints. Possible cause - differences in column
temperature gradients. Data is suspect. Updated calibration coefficients are needed to derive
correct supersaturation values from the recorded temperatures. |
Suggestions: | Supersaturation values need to be recalculated. |
Measurements: | sgpaosccn200S01.a1: - AOS CCN number concentration by bin, second CCN(N_CCN_dN_2)
- AOS CCN droplet count by bin size, first column(N_CCN_dN_1)
- AOS number concentration of CCN at supersaturation step 1(N_CCN_1)
- AOS CCN sample supersaturation set point value reported by instrument, first
column(CCN_supersaturation_set_point_1) - AOS number concentration of CCN at supersaturation step 2(N_CCN_2)
- AOS CCN sample supersaturation set point value reported by instrument, second
column(CCN_supersaturation_set_point_2) more
sgpaosccn200S01.00: - (Raw data stream - documentation not supported)
more
|