DQRID : D160617.16
Start DateStart TimeEnd DateEnd Time
04/04/2016180005/12/20160429
more
Subject:
AWR/SKYRAD/M1 - Heater not operating
DataStreams:awrskyrad60sM1.b1, awrskyrad20sM1.a0
Description:
The IRT did not measure sky temperature correctly when the ambient
the temperature was below approximately -25C, because the IRT heater was not operating, 
and due to frost depositing on the mirror and subsequent cleaning with warm ethanol.
Suggestions: 
Use data with caution
Measurements:awrskyrad60sM1.b1:
  • Sky Infra-Red Temperature Minima(sky_ir_temp_min)
  • Sky infrared temperature, standard deviation(sky_ir_temp_std)
  • Sky Infra Red Temperature Maxima(sky_ir_temp_max)
  • Sky Infra-Red Temperature(sky_ir_temp)
more
awrskyrad20sM1.a0:
  • Instantaneous Sky/Cloud Infrared(inst_sky_ir_signal)
more