DQRID : D160707.1
Start DateStart TimeEnd DateEnd Time
05/12/2016090305/17/20161618
more
Subject:
ASI/TRACEGAS/M1 - NOy Channel failed
DataStreams:asiaosnoxM1.a1
Description:
Instrument Mass Flow Controller for the NOy channel died (slowly).  It was replaced and 
normal function restored.
Suggestions: 
No work around
Measurements:asiaosnoxM1.a1:
  • One minute average of NOy values using nominal sensitivity correction and
    nominal Mo converter efficiency(noy)
more