Subject:
ASI/AETH/M1 - Negative black carbon mass
DataStreams:asiaosaethM1.00, asiaosaeth1spotM1.a1
Description:
Due to a design flaw - separate plumbing lines on the sample and reference lines - in a 
small laboratory volume differences in the RH between these two lines will oscillate when 
the laboratory A/C cycles on and off. This is a known problem with the Aethelometer model 
(AE-31) currently deployed in the MAOS.