Subject:
ENA/OZONE/C1 - Water condensing in the line
DataStreams:enaaosozoneC1.01, enaaosozoneC1.a1, enaaosozoneC1.00
Description:
Instrument suffered major oscillations in the signal.  Working with Operator it was 
discovered that the Split System Air Conditioner was blowing directly on the Ozone detector.  
Water was condensing in the line.  The air was redirected and the temperature setpoint 
raised.  This reduced but didn't entirely eliminate the oscillations.
Suggestions: 
Do not use data