Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/03/2016 | 1655 | 07/21/2016 | 2359 |
Subject: | ASI/SASZE/M1 - Spectrometer chiller temperature variations |
DataStreams: | asisaszefilterbandsM1.a1, asisaszevisM1.a1 |
Description: | Spectrometer chiller temperature variations caused by an incorrectly placed sensor are apparent as correlated modulations in the signal from the UV-VIS CCD spectrometer. The magnitude of signal oscillation is small and is only apparent under clear sky for wavelengths < 450 nm. The variations were reduced to negligible levels by repositioning the sensor on Aug. 21, 2016. |
Suggestions: | null |
Measurements: | asisaszefilterbandsM1.a1: more
asisaszevisM1.a1: more |