Subject:
OLI/RL/M1 - Photon counting and analog signals  contaminated with ambient and laser light
DataStreams:olirlM1.a0
Description:
A malfunction of the narrow-field-of-view (NFOV) "flipper" caused light to leak into the 
mode 0 measurement. The mode 0 measurement is used to characterize the response of the 
system in the absence of any light (i.e. its used to measure the noise). The measurement is 
done once per hour. During the period in question the mechanism that's used to block the 
aperture (flipper) for the NFOV channels malfunctioned.
A new flipper was installed on 2016-09-21.
Suggestions: 
null
Measurements:olirlM1.a0:
  • Summed analog signal in the high water vapor channel(water_analog_high)
  • Summed analog signal in the high elastic channel(elastic_analog_high)
  • Number of photons counted in the high nitrogen channel(nitrogen_counts_high)
  • Count rate in the high water vapor channel(water_counts_high)
  • Number of photons counted in the high elastic (aerosol) channel(elastic_counts_high)
  • Summed analog signal in the high nitrogen channel(nitrogen_analog_high)
  • Summed analog signal in the high depolarization channel(depolarization_analog_high)
  • Number of photons counted in the temperature 2 channel(t2_counts_high)
  • Summed analog signal in the temperature 1 channel(t1_analog_high)
  • Wavelength of each of the 6 filters measured by the MFR radiometers(filter)
  • Summed analog signal in the temperature 2 channel(t2_analog_high)
  • Number of photons counted in the temperature 1 channel(t1_counts_high)
  • Count rate in the high depolarization channel(depolarization_counts_high)
more

Close this window