DQRID : D951020.1
Start DateStart TimeEnd DateEnd Time
10/05/1995153010/05/19951830
more
Subject:
SGP/EBBR/E9 - 3-month Check
DataStreams:sgp5ebbrE9.b1, sgp30ebbrE9.b1, sgp15ebbrE9.b1
Description:
During EBBR 3 month checks the Automatic Exchange Mechanism is disabled
(reflected by home signals near zero) so that the two aspirator units can be
placed at the same height.  The measurements during this period, at least,
do not produce correct outputs of bowen ratio (bowen), sensible heat flux (h),
or latent heat flux (e), nor do the air temperature and relative humidity
measurements indicate gradients.  Checks performed before or after the AEM
is disabled also lead to some incorrect data.
Suggestions: 
Measurements:sgp30ebbrE9.b1:
  • Bottom air temperature(temp_air_bottom)
  • Top air temperature(temp_air_top)
  • Bottom relative humidity (fractional)(rh_bottom_fraction)
  • Bottom vapor pressure(vapor_pressure_bottom)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
  • Top vapor pressure(vapor_pressure_top)
  • Top T/RH sensor temperature(temp_trh_top)
  • Surface latent heat flux(latent_heat_flux)
  • Top relative humidity (fractional)(rh_top_fraction)
  • Bowen ratio(bowen_ratio)
  • Surface sensible heat flux(sensible_heat_flux)
more
sgp15ebbrE9.b1:
  • Left air temperature(temp_air_left)
  • Left relative humidity (mV)(rh_left_mv)
  • Right T/RH sensor temperature (resistance ratio)(temp_trh_right_rr)
  • Right air temperature(temp_air_right)
  • Right relative humidity (mV)(rh_right_mv)
  • Left T/RH sensor temperature (resistance ratio)(temp_trh_left_rr)
more
sgp5ebbrE9.b1:
  • Bottom relative humidity (fractional)(rh_bottom_fraction)
  • Bottom air temperature(temp_air_bottom)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
  • Top vapor pressure(vapor_pressure_top)
  • Top T/RH sensor temperature(temp_trh_top)
  • Top air temperature(temp_air_top)
  • Top relative humidity (fractional)(rh_top_fraction)
  • Bottom vapor pressure(vapor_pressure_bottom)
more