DQRID : D970925.1
Start DateStart TimeEnd DateEnd Time
09/02/1997200409/04/19972057
more
Subject:
SGP/TWR/C1 - 25m Low battery voltage affected data
DataStreams:sgp1twr25mC1.a0, sgp30twr25mC1.a1
Description:
Shortly after 2003 GMT on 2 September 1997 a GFCI circuit breaker tripped in the tower 
base AC box, resulting in a drop in CR21X voltage of 0.4 volts in seven minutes.  This 
resulted in a steady drop in battery voltage for the next 30 hours, and incorrect, out of 
range, or missing values for temp, rh, and vap_pres.
Suggestions: 
Measurements:sgp1twr25mC1.a0:
  • -180.0 - +180.0(alt)
  • lon(lon)
  • base time in epoch(base_time)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • Aspirator flow status (% of time with proper flow)(aspirator)
  • North Latitude(lat)
  • Environmental Temperature (hybrid-b grid)(temp)
  • Time offset from base_time(time_offset)
  • Relative Humidity (hybrid-b grid)(rh)
  • Average battery voltage(vbat)
more
sgp30twr25mC1.a1:
  • Standard Deviation of Air Temperature(sd_temp)
  • Relative Humidity (hybrid-b grid)(rh)
  • Vapor Pressure (kiloPascals)(vap_pres)
  • base time in epoch(base_time)
  • lon(lon)
  • -180.0 - +180.0(alt)
  • Standard deviation of vapor pressure(sd_vap_pres)
  • Environmental Temperature (hybrid-b grid)(temp)
  • Aspirator flow status (% of time with proper flow)(aspirator)
  • North Latitude(lat)
  • Standard Deviation of Relative Humidity(sd_rh)
  • Time offset from base_time(time_offset)
more

Close this window