Subject: | SGP/SONDE/C1 - Reprocess: Incorrect surface press and rh values |
DataStreams: | sgpsondewrpnC1.a1, sgpsondewrpnC1.a0
|
Description: | Site ops noted that incorrect values for surface pressure and relative humidity were
entered for the subject sounding. Pressure was entered as 986.1 instead of 981.1, and RH was
entered as 27 instead of 57. The incorrect surface pressure entry will cause the computed
altitudes to be incorrect for the sounding and the ascent rates to be incorrect for the
first 30 seconds or so. The incorrect surface RH will not affect any of the subsequent
measurements. |
Suggestions: | Wait for reprocessed data |
Measurements: | sgpsondewrpnC1.a0: - Dry Bulb Temperature(tdry)
- Ascent rate(asc)
- Dewpoint Temperature(dp)
- -180.0 - +180.0(alt)
- Relative Humidity (hybrid-b grid)(rh)
- Retrieved pressure profile(pres)
more
sgpsondewrpnC1.a1: - Ascent rate(asc)
- Relative Humidity (hybrid-b grid)(rh)
- -180.0 - +180.0(alt)
- Dewpoint Temperature(dp)
- Dry Bulb Temperature(tdry)
- Retrieved pressure profile(pres)
more
|