DQRID : D990308.2
Start DateStart TimeEnd DateEnd Time
11/06/1997000011/26/19971720
more
Subject:
SGP/SIRS/E2 - Reprocess: wrong calibration
DataStreams:sgpsirsE2.01, sgpsirsE2.a1
Description:
Mike Splitt found the following in the Site Ops OMIS data base from E2 on
11/26/97 :


        Found that the wrong calibration sticker was placed on the nip
        and the wrong Cal factor had been entered in the program. Also
        while editing the program I found that I had the wrong Cal
        factor for the DS. I inserted the correct Cal factors for the DS
        and Nip, complied and down loaded the new program into
        SIRS_02 at 1720GMT on 11/26/97

This entry was made by Craig Webb.

The problem was reported in P971121.1
Suggestions: 
These data need to be reprocessed.
Measurements:sgpsirsE2.01:
  • (Raw data stream - documentation not supported)
more
sgpsirsE2.a1:
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Standard
    Deviation(down_short_hemisp_std)
  • Downwelling shortwave hemispheric irradiance, pyranometer, maxima(down_short_hemisp_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Down-welling unshaded pyranometer voltage(down_short_hemisp)
  • Shortwave Direct Normal Irradiance, Pyrgeometer, Standard Deviation(short_direct_normal_std)
  • Downwelling Shortwave Hemispheric Irradiance, Pyranometer, Minima(down_short_hemisp_min)
  • Shortwave direct normal irradiance, pyrheliometer, minima(short_direct_normal_min)
  • Shortwave direct normal irradiance, pyrheliometer(short_direct_normal)
more