DQRID : D990311.1
Start DateStart TimeEnd DateEnd Time
02/02/1999040002/23/19991500
more
Subject:
SGP/EBBR/E15 - Temperatures Incorrect
DataStreams:sgp5ebbrE15.b1, sgp30ebbrE15.b1, sgp15ebbrE15.b1
Description:
Poor contact of the reference temperature probe leads to the J-panel or
multiplexer caused frequent offscale and incorrect values of the reference
temperature.  All other temperature measurements (air, soil, and humidity 
probe) depend on the reference temperature for a correction.  Therefore all
temperature measurements are frequently offscale and incorrect during the 
specified period.  The data clearly shows the offscale times.
Suggestions: 
Measurements:sgp5ebbrE15.b1:
  • Top T/RH sensor temperature(temp_trh_top)
  • Top air temperature(temp_air_top)
  • Bottom air temperature(temp_air_bottom)
  • Reference temperature in enclosure(temp_reference)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
more
sgp15ebbrE15.b1:
  • Left air temperature(temp_air_left)
  • Reference temperature in enclosure (resistance ratio)(temp_reference_rr)
  • Soil temperature 4 (resistance ratio)(soil_temp_4_rr)
  • Soil temperature 2 (resistance ratio)(soil_temp_2_rr)
  • Right T/RH sensor temperature (resistance ratio)(temp_trh_right_rr)
  • Soil temperature 5 (resistance ratio)(soil_temp_5_rr)
  • Soil temperature 3 (resistance ratio)(soil_temp_3_rr)
  • Soil temperature 1 (resistance ratio)(soil_temp_1_rr)
  • Right air temperature(temp_air_right)
  • Left T/RH sensor temperature (resistance ratio)(temp_trh_left_rr)
more
sgp30ebbrE15.b1:
  • Bottom air temperature(temp_air_bottom)
  • Half hour averaged soil temperature 4(soil_temp_4)
  • Soil temperature mean 3(soil_temp_3)
  • Soil temperature 2(soil_temp_2)
  • Soil temperature 1(soil_temp_1)
  • Half hour averaged soil temperature 5(soil_temp_5)
  • Reference temperature in enclosure(temp_reference)
  • Bottom T/RH sensor temperature(temp_trh_bottom)
  • Top air temperature(temp_air_top)
  • Top T/RH sensor temperature(temp_trh_top)
more