DQRID : D990614.1
Start DateStart TimeEnd DateEnd Time
05/21/1999141305/27/19991819
more
Subject:
SGP/AOS/C1 - Invalid absorption data
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
During the period noted, a valve position error with the 1-um/10-um impactor switching 
valve occurred.  The result of this error was to invalidate all switched data.  The software 
correctly inserted missing value codes in the nephelometer data, but the PSAP light 
absorption photometer for some reason reported a constant value over this period.
Suggestions: 
Do not use PSAP data during this interval.
Measurements:sgpaosauxC1.a0:
  • CLAP reference signal(BapRef)
  • CLAP Signal(BapSignal)
  • Transmittance through CLAP filter(BapFiltr)
  • CLAP flow(BapFlow)