Subject:
SGP/THWAPS - RH values low by factor of 1.1
DataStreams:DsgpthwapsC1.b1, sgpthwapsC1.b1, DsgpthwapsC1.a1, sgpthwapsC1.00, sgpthwapsC1.a1
Description:
Because of instrument mentor stupidity, all C1 THWAPS RH data collected
from the begin date of this DQR are low by a factor of 1.1.  This error
also will affect the vapor pressure data.  The problem is that the CSI
data logger is expecting the data to be scaled 0-100%RH on the range 0-1V,
but it is now being scaled 0-110%RH on the range 0-1V.
Suggestions: 
The problem can be fixed by reprogramming the CSI data
logger.I will be sending instructions for doing so down to Site Ops soon.
The same problem affects the B1 THWAPS.  I will be filing
a separate DQR for that.  Data users can obtaine the correct
RH by multiplying the values in the file by 1.1.