Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/13/1999 | 1420 | 10/14/1999 | 2230 |
Subject: | SGP/THWAPS - THWAPS pressure too high by factor of 1.1 |
DataStreams: | DsgpthwapsC1.b1, DsgpthwapsC1.00, DsgpthwapsC1.a1 |
Description: | When attempting to reprogram the C1 THWAPS Campbell Scientific data logger to correct the problem (previously reported) with RH, Site Ops mistakenly changed the scaling of the pressure sensor instead. Pressure data in the ingested file is 1.1 too high. The RH is 1.1 too low (subject of a different DQR). |
Suggestions: | Users may divide the listed pressure by 1.1 to obtain an estimate of the correct value. The problem has been fixed . |
Measurements: | DsgpthwapsC1.b1: more
DsgpthwapsC1.00: more DsgpthwapsC1.a1: more |