Data Quality Reports for Session: 101233 User: chollow Completed: 08/22/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040127.3TWP/SONDE/C2 - Broken temperature sensorstwpsondewnpnC2.b1
D040421.1TWP/SMET/C2 - Wind Sensor failuretwpsmet60sC2.b1
D040425.1TWP/SONDE/C2 - Bad RH values alofttwpsondewnpnC2.b1
D040711.2TWP/SONDE/C2 - RH sensor failure near surfacetwpsondewnpnC2.b1
D040722.6TWP/SMET/C2 - All data missing due to logger changetwpsmet60sC2.b1
D040722.7TWP/SMET/C2 - connection and timing issues with new loggertwpsmet60sC2.b1
D040726.1TWP/SMET/C2 - Negative rainfall ratestwpsmet60sC2.b1
D040815.1TWP/SONDE/C2 - Operator used wrong calibration tapetwpsondewnpnC2.b1
D040824.2TWP/SMET/C2 - Incorrect rainrate equation used.twpsmet60sC2.b1
D040921.1TWP/SONDE/C2 - RH sensor failure below 9kmtwpsondewnpnC2.b1
D040921.2TWP/SONDE/C2 - Bad RH values alofttwpsondewnpnC2.b1
D050301.1TWP/SMET/C2 - Instrument failuretwpsmet60sC2.b1
D050322.2TWP/SMET/C2 - Upper wind sensor failuretwpsmet60sC2.b1
D050601.5TWP/SMET/C2 - Reprocess: Incorrect Units Listed for Vapor Pressuretwpsmet60sC2.b1
D050714.1TWP/SMET/C2 - RH failuretwpsmet60sC2.b1
D050913.3TWP/SONDE/C2 - Bad RH sensortwpsondewnpnC2.b1
D050920.4TWP/SMET/C2 - Instrument maintenancetwpsmet60sC2.b1
D050921.2TWP/SMET/C2 - ORG data incorrecttwpsmet60sC2.b1
D051005.6TWP/SMET/C2 - Tower down for calibration activitiestwpsmet60sC2.b1
D051122.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D051219.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D060308.1TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D060403.2TWP/SONDE/C2 - Broken temperature sensortwpsondewnpnC2.b1
D060418.1TWP/SONDE/C2 - Bad RH in lower 8kmtwpsondewnpnC2.b1


DQRID : D040127.3
Start DateStart TimeEnd DateEnd Time
10/13/2003004010/13/20030203
11/08/2003233511/09/20030215
11/18/2003120511/18/20031340
11/19/2003115611/19/20031346
11/21/2003004411/21/20030113
12/20/2003122512/20/20031408
12/21/2003005612/21/20030126
12/21/2003115512/21/20031344
12/23/2003004212/23/20030212
12/23/2003113712/23/20031259
12/23/2003234212/24/20030132
01/08/2004113501/08/20041309
01/09/2004115701/09/20041342
01/09/2004135901/09/20041537
01/12/2004234201/13/20040214
01/16/2004140901/16/20041548
01/16/2004233701/17/20040109
01/18/2004113501/18/20041256
01/19/2004022301/19/20040346
01/19/2004232901/20/20040157
01/21/2004125201/21/20041400
01/22/2004005201/22/20040227
01/23/2004113701/23/20041319
01/26/2004120201/26/20040146
02/17/2004140602/17/20041536
02/26/2004122102/26/20041350
03/09/2004114503/09/20041315
03/14/2004113103/14/20041300
03/22/2004113403/22/20041300
03/29/2004233203/30/20040047
03/30/2004233403/31/20040059
04/11/2004234804/12/20040120
05/01/2004233505/02/20040110
05/02/2004233605/03/20040109
05/13/2004233405/14/20040112
05/14/2004140605/14/20041411
06/07/2004022406/07/20040351
06/25/2004114606/25/20041328
07/02/2004005907/02/20040217
07/20/2004022007/20/20040358
07/24/2004123507/24/20041355
08/22/2004114408/22/20041207
09/08/2004142509/08/20041605
09/08/2004233809/08/20042345
10/18/2004234610/19/20040128
11/29/2004232111/30/20040105
12/08/2004114312/08/20041306
12/30/2004113012/30/20041239
Subject:
TWP/SONDE/C2 - Broken temperature sensors
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensors on these radiosondes broke after launch.  All
temperature-related data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Dry-Bulb Temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040421.1
Start DateStart TimeEnd DateEnd Time
04/10/2004000004/24/20042303
Subject:
TWP/SMET/C2 - Wind Sensor failure
DataStreams:twpsmet60sC2.b1
Description:
Wind sensor #2 (lower one) failed.  Wind direction showed little or
no variation, unlike the colocated sensor #1.  This caused lower
wind speed values as the sensor was not rotating properly to orient
itself parallel to the wind stream.  The sensor was replaced.
Measurements:twpsmet60sC2.b1:
  • Wind #2 speed maximum(wind2_spd_max)
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • Wind #2 direction standard deviation(wind2_dir_sd)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)
  • Wind #2 direction vector average(wind2_dir_vec_avg)
  • Wind #2 speed standard deviation(wind2_spd_sd)
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • Wind #2 speed minimum(wind2_spd_min)


Back To Table of Contents

DQRID : D040425.1
Start DateStart TimeEnd DateEnd Time
04/21/2004123204/21/20041248
Subject:
TWP/SONDE/C2 - Bad RH values aloft
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit on this sonde failed about 24 minutes into 
the flight.  The RH values look reasonable, though somewhat noisy, from the surface to 
approximately 6.9 km (439 hPa) at which point they begin to oscillate between 0 and 100 %.  
This behavior stops when the sonde reaches approximately 11.5km (232 hPa) when the air 
temperature drops below -40 degC.  The oscillation stops at that point, though the RH 
remains higher than might be expected even under supersaturated conditions.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040711.2
Start DateStart TimeEnd DateEnd Time
07/10/2004123807/10/20041315
Subject:
TWP/SONDE/C2  - RH sensor failure near surface
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040722.6
Start DateStart TimeEnd DateEnd Time
07/06/2004200307/07/20040137
Subject:
TWP/SMET/C2 - All data missing due to logger change
DataStreams:twpsmet60sC2.b1
Description:
All data are missing at this site during this time period due to collection system, logger 
and enclosure change.  Data resumed after change.
Measurements:twpsmet60sC2.b1:
  • Relative humidity mean(relh_mean)
  • Wind #1 direction standard deviation(wind1_dir_sd)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Wind #1 speed maximum(wind1_spd_max)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)
  • Wind #1 speed standard deviation(wind1_spd_sd)
  • Time offset from base_time(base_time)
  • Max Temperature at lowest Model Level- 1 hr(temp_max)
  • east longitude for all the input platforms.(lon)
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • Internal Logger Temperature(logger_temp)
  • Bit data. See details.(bit_data)
  • Time offset of tweaks from base_time(time_offset)
  • Wind #2 direction standard deviation(wind2_dir_sd)
  • Wind #1 speed arithmetic average(wind1_spd_arith_avg)
  • Wind #1 speed vector average(wind1_spd_vec_avg)
  • precipitation minimum(precip_min)
  • Wind #2 speed minimum(wind2_spd_min)
  • Vapor pressure mean(vappress_mean)
  • Internal voltage(internal_voltage)
  • Std Dev of Relative Humidity or Hardware Error Code(relh_sd)
  • Wind #2 speed maximum(wind2_spd_max)
  • Wind #1 speed minimum(wind1_spd_min)
  • GPS raw data filename(filename)
  • Temperature standard deviation(temp_sd)
  • Min Temperature at lowest Model Level- 1 hr(temp_min)
  • Atmospheric pressure(atmos_pressure)
  • altitude above sea levelaltunits(alt)
  • Relative humidity minimum(relh_min)
  • Wind #2 speed standard deviation(wind2_spd_sd)
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • Vapor pressure standard deviation(vappress_sd)
  • north latitude for all the input platforms.(lat)
  • Relative humidity maximum(relh_max)
  • Precipitation maximum(precip_max)
  • Wind #1 direction vector average(wind1_dir_vec_avg)
  • External voltage(external_voltage)
  • Wind #2 direction vector average(wind2_dir_vec_avg)
  • Wind #1 speed maximum gust(wind1_spd_max_gust)
  • Precipitation mean(precip_mean)
  • Precipitation standard deviation(precip_sd)


Back To Table of Contents

DQRID : D040722.7
Start DateStart TimeEnd DateEnd Time
07/07/2004013807/07/20042018
Subject:
TWP/SMET/C2 - connection and timing issues with new logger
DataStreams:twpsmet60sC2.b1
Description:
Atmospheric pressure data are intermittent during this time period due to connection 
problems and program timing issues with new logger installation.  Data resumed normal 
operations after this time period.
Measurements:twpsmet60sC2.b1:
  • Atmospheric pressure(atmos_pressure)


Back To Table of Contents

DQRID : D040726.1
Start DateStart TimeEnd DateEnd Time
10/21/1998031107/06/20040138
Subject:
TWP/SMET/C2 - Negative rainfall rates
DataStreams:twpsmet60sC2.b1
Description:
Due to limitations of the Zeno dataloggers all voltage measurements
from the optical rain gauge (ORG) were converted to rainfall rates. 
This resulted in negative values for rain rates.  Any value below
0.10 mm/hr should be considered to be 0 mm/hr.  New dataloggers 
alleviated this problem.
Measurements:twpsmet60sC2.b1:
  • Precipitation maximum(precip_max)
  • Precipitation mean(precip_mean)
  • precipitation minimum(precip_min)
  • Precipitation standard deviation(precip_sd)


Back To Table of Contents

DQRID : D040815.1
Start DateStart TimeEnd DateEnd Time
08/12/2004004408/12/20040233
Subject:
TWP/SONDE/C2 - Operator used wrong calibration tape
DataStreams:twpsondewnpnC2.b1
Description:
It appears as though the opperator used the wrong calibration tape for this sounding.  The 
primary symptom is a jump in temperature from a surface value of 30.1 degC to the first 
value aloft of 61.0 degC.  Secondly, the recorded serial number for this sounding is 
Z1430179 which is the same serial number recorded for the previous sounding.

This problem points to three failures of procedure (I have notified TWP site ops by 
e-mail).  The first failure is using the wrong tape. The second failure is not checking the 
radiosonde values before launch which relates to the third failure of not entering the 
radiosonde values as the surface values of the sounding.
Measurements:twpsondewnpnC2.b1:
  • Dry-Bulb Temperature(tdry)
  • Ascent Rate(asc)
  • east longitude for all the input platforms.(lon)
  • north latitude for all the input platforms.(lat)
  • Wind Status(wstat)
  • surface wind speed(wspd)
  • model meridional wind velocity(v_wind)
  • pressure at constant pressure surface(pres)
  • Time offset of tweaks from base_time(time_offset)
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Surface dew point temperature(dp)
  • Time offset from base_time(base_time)
  • model zonal wind velocity(u_wind)
  • Wind Direction(deg)
  • altitude above sea levelaltunits(alt)


Back To Table of Contents

DQRID : D040824.2
Start DateStart TimeEnd DateEnd Time
07/07/2004013708/17/20042157
Subject:
TWP/SMET/C2 - Incorrect rainrate equation used.
DataStreams:twpsmet60sC2.b1
Description:
There are two different Optical Raingages in use at the TWP sites.  The incorrect equation 
was used at this site.  The incorrect equation is (25*(V^1.87)) - 0.15.
The correct equation is (20*(V^2)) - 0.05.  The correct equation was installed.
Measurements:twpsmet60sC2.b1:
  • Precipitation maximum(precip_max)
  • Precipitation mean(precip_mean)
  • precipitation minimum(precip_min)
  • Precipitation standard deviation(precip_sd)


Back To Table of Contents

DQRID : D040921.1
Start DateStart TimeEnd DateEnd Time
09/16/2004113509/16/20041210
Subject:
TWP/SONDE/C2  - RH sensor failure below 9km
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D040921.2
Start DateStart TimeEnd DateEnd Time
09/17/2004114509/17/20041216
09/18/2004005809/18/20040133
09/19/2004140809/19/20041447
Subject:
TWP/SONDE/C2 - Bad RH values aloft
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit on these sondes failed after launch but 
resume proper operation when the ambient temperature dropped below ~40 degC.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D050301.1
Start DateStart TimeEnd DateEnd Time
02/12/2005195503/03/20050430
Subject:
TWP/SMET/C2 - Instrument failure
DataStreams:twpsmet60sC2.b1
Description:
Upper wind sensor failed. Wind direction was reported at a near constant value of 69 
degrees and was significantly different than the lower wind sensor's direction values. Sensor 
replaced.
Measurements:twpsmet60sC2.b1:
  • Wind #1 direction standard deviation(wind1_dir_sd)
  • Wind #1 direction vector average(wind1_dir_vec_avg)


Back To Table of Contents

DQRID : D050322.2
Start DateStart TimeEnd DateEnd Time
03/09/2005230004/17/20050332
Subject:
TWP/SMET/C2 - Upper wind sensor failure
DataStreams:twpsmet60sC2.b1
Description:
After replacement on 03/03/2005 the upper wind sensor began reporting erroneous values 
again on 03/09/2005 becoming very apparent after 03/17/2005.  Wind speed data appears 
reasonable but the wind direction data do not agree well with the lower wind sensor.
Measurements:twpsmet60sC2.b1:
  • Wind #1 direction standard deviation(wind1_dir_sd)
  • Wind #1 direction vector average(wind1_dir_vec_avg)


Back To Table of Contents

DQRID : D050601.5
Start DateStart TimeEnd DateEnd Time
07/07/2004000005/19/20052359
Subject:
TWP/SMET/C2 - Reprocess: Incorrect Units Listed for Vapor Pressure
DataStreams:twpsmet60sC2.b1
Description:
The netcdf header file incorrectly listed the units of Vapor Pressure as hPa when in fact 
they were kPa.
Measurements:twpsmet60sC2.b1:
  • Vapor pressure standard deviation(vappress_sd)
  • Vapor pressure mean(vappress_mean)


Back To Table of Contents

DQRID : D050714.1
Start DateStart TimeEnd DateEnd Time
07/10/2005190007/14/20052226
Subject:
TWP/SMET/C2 - RH failure
DataStreams:twpsmet60sC2.b1
Description:
The RH values were frequently over MAX value of 104% starting on 7/10/2005 through end of 
period.  Sensor was replaced.
Measurements:twpsmet60sC2.b1:
  • Relative humidity mean(relh_mean)
  • Relative humidity maximum(relh_max)
  • Std Dev of Relative Humidity or Hardware Error Code(relh_sd)
  • Relative humidity minimum(relh_min)


Back To Table of Contents

DQRID : D050913.3
Start DateStart TimeEnd DateEnd Time
07/03/2005235007/04/20050024
07/10/2005234007/11/20050021
07/19/2005233207/20/20050016
Subject:
TWP/SONDE/C2 - Bad RH sensor
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below -50 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D050920.4
Start DateStart TimeEnd DateEnd Time
09/21/2005162509/21/20051815
Subject:
TWP/SMET/C2 - Instrument maintenance
DataStreams:twpsmet60sC2.b1
Description:
Instrument calibration and testing were taking place for brief periods.
Measurements:twpsmet60sC2.b1:
  • Relative humidity mean(relh_mean)
  • Wind #1 direction standard deviation(wind1_dir_sd)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Wind #1 speed maximum(wind1_spd_max)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)
  • Wind #1 speed standard deviation(wind1_spd_sd)
  • Max Temperature at lowest Model Level- 1 hr(temp_max)
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • Wind #2 direction standard deviation(wind2_dir_sd)
  • Wind #1 speed vector average(wind1_spd_vec_avg)
  • Wind #1 speed arithmetic average(wind1_spd_arith_avg)
  • Vapor pressure mean(vappress_mean)
  • Wind #2 speed minimum(wind2_spd_min)
  • Std Dev of Relative Humidity or Hardware Error Code(relh_sd)
  • Wind #1 speed minimum(wind1_spd_min)
  • Wind #2 speed maximum(wind2_spd_max)
  • Temperature standard deviation(temp_sd)
  • Min Temperature at lowest Model Level- 1 hr(temp_min)
  • Wind #2 speed standard deviation(wind2_spd_sd)
  • Relative humidity minimum(relh_min)
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • Vapor pressure standard deviation(vappress_sd)
  • Relative humidity maximum(relh_max)
  • Wind #1 direction vector average(wind1_dir_vec_avg)
  • Wind #2 direction vector average(wind2_dir_vec_avg)
  • Wind #1 speed maximum gust(wind1_spd_max_gust)


Back To Table of Contents

DQRID : D050921.2
Start DateStart TimeEnd DateEnd Time
09/19/2005080909/19/20051902
Subject:
TWP/SMET/C2 - ORG data incorrect
DataStreams:twpsmet60sC2.b1
Description:
Data from the ORG was incorrect due to heavy condensation on the lens.
Measurements:twpsmet60sC2.b1:
  • Precipitation maximum(precip_max)
  • Precipitation mean(precip_mean)
  • precipitation minimum(precip_min)
  • Precipitation standard deviation(precip_sd)


Back To Table of Contents

DQRID : D051005.6
Start DateStart TimeEnd DateEnd Time
09/19/2005020009/19/20050330
Subject:
TWP/SMET/C2 - Tower down for calibration activities
DataStreams:twpsmet60sC2.b1
Description:
Tower lowered, wind vanes spun at different frequencies, and ORG tested with electronic 
rain.
Measurements:twpsmet60sC2.b1:
  • Relative humidity mean(relh_mean)
  • Wind #1 direction standard deviation(wind1_dir_sd)
  • Mean Air Temperature or Hardware Error(temp_mean)
  • Wind #1 speed maximum(wind1_spd_max)
  • Wind #2 speed maximum gust(wind2_spd_max_gust)
  • Wind #1 speed standard deviation(wind1_spd_sd)
  • Max Temperature at lowest Model Level- 1 hr(temp_max)
  • Wind #2 speed vector average(wind2_spd_vec_avg)
  • Wind #2 direction standard deviation(wind2_dir_sd)
  • Wind #1 speed arithmetic average(wind1_spd_arith_avg)
  • Wind #1 speed vector average(wind1_spd_vec_avg)
  • precipitation minimum(precip_min)
  • Vapor pressure mean(vappress_mean)
  • Wind #2 speed minimum(wind2_spd_min)
  • Std Dev of Relative Humidity or Hardware Error Code(relh_sd)
  • Wind #1 speed minimum(wind1_spd_min)
  • Wind #2 speed maximum(wind2_spd_max)
  • Temperature standard deviation(temp_sd)
  • Min Temperature at lowest Model Level- 1 hr(temp_min)
  • Relative humidity minimum(relh_min)
  • Wind #2 speed standard deviation(wind2_spd_sd)
  • Wind #2 speed arithmetic average(wind2_spd_arith_avg)
  • Vapor pressure standard deviation(vappress_sd)
  • Precipitation maximum(precip_max)
  • Relative humidity maximum(relh_max)
  • Wind #1 direction vector average(wind1_dir_vec_avg)
  • Wind #2 direction vector average(wind2_dir_vec_avg)
  • Wind #1 speed maximum gust(wind1_spd_max_gust)
  • Precipitation mean(precip_mean)
  • Precipitation standard deviation(precip_sd)


Back To Table of Contents

DQRID : D051122.2
Start DateStart TimeEnd DateEnd Time
11/19/2005233411/20/20050153
11/21/2005233211/22/20050140
Subject:
TWP/SONDE/C2 - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Dry-Bulb Temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D051219.2
Start DateStart TimeEnd DateEnd Time
12/07/2005115312/07/20051323
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Dry-Bulb Temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D060308.1
Start DateStart TimeEnd DateEnd Time
01/15/2006114401/15/20061328
01/19/2006131801/19/20061320
01/22/2006114901/22/20061315
02/14/2006005502/14/20060207
02/21/2006023202/21/20060359
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Dry-Bulb Temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D060403.2
Start DateStart TimeEnd DateEnd Time
03/27/2006004303/27/20060208
Subject:
TWP/SONDE/C2  - Broken temperature sensor
DataStreams:twpsondewnpnC2.b1
Description:
The temperature sensor on this radiosonde broke after launch.  All the temperature-related 
data are incorrect.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Dry-Bulb Temperature(tdry)
  • Surface dew point temperature(dp)


Back To Table of Contents

DQRID : D060418.1
Start DateStart TimeEnd DateEnd Time
04/10/2006024704/10/20060314
Subject:
TWP/SONDE/C2  - Bad RH in lower 8km
DataStreams:twpsondewnpnC2.b1
Description:
It appears that the RH sensor heating circuit failed during this sounding.  The RH 
oscillates rapidly between low (~0 %RH) and not so low (ambient %RH) and the oscillation seems 
to stop when the temperature is below ~35 degC.  According to Vaisala these are symptoms 
of failure in the ASIC (application specific integrated circuit) that controls the heating.
Measurements:twpsondewnpnC2.b1:
  • Relative humidity scaled, by total column amount from MWR(rh)
  • Surface dew point temperature(dp)


Back To Table of Contents



END OF DATA