DQR ID | Subject | Data Streams Affected |
---|---|---|
D021211.1 | TWP/MET/C3 - WIND DIRECTION SENSOR #2 MISALIGNED | twpsmet60sC3.a1, twpsmet60sC3.b1 |
D030212.1 | TWP/SMET/C3 - Intermittent Pressure Sensor Failure | twpsmet60sC3.a1, twpsmet60sC3.b1 |
D031103.2 | TWP/SMET/C3 - Data reported at 15sec rather than 60s intervals | twpsmet60sC3.a1, twpsmet60sC3.b1 |
D040722.4 | TWP/SMET/C3 - Rain blocked pressure port | twpsmet60sC3.b1 |
D040722.5 | TWP/SMET/C3 - Data missing due to Logger Change | twpsmet60sC3.b1 |
D040726.3 | TWP/SMET/C3 - Negative rainfall rates | twpsmet60sC3.b1 |
D050210.1 | TWP/SMET/C3 - Data non-representative | twpsmet60sC3.b1 |
D050418.2 | TWP/SMET/C3 - Instrument logger problem, Sensor failure | twpsmet60sC3.b1 |
D050601.3 | TWP/SMET/C3 - Reprocess: Incorrect Units Listed for Vapor Pressure | twpsmet60sC3.b1 |
D050808.3 | TWP/SMET/C3 - Data not representative | twpsmet60sC3.b1 |
D050831.4 | TWP/SMET/C3 - Towrr lowered for sensor comparisons, data not representative | twpsmet60sC3.b1 |
D051115.1 | TWP/SMET/C3 - Upper wind sensor failure | twpsmet60sC3.b1 |
D051117.2 | TWP/SMET/C3 - RH Values Artificially High | twpsmet60sC3.b1 |
D051117.3 | TWP/SMET/C3 - Instrument failure | twpsmet60sC3.b1 |
D060502.3 | TWP/SMET/C3 - Reprocess: Windspeed2 calibration error | twpsmet60sC3.b1 |
D060627.1 | TWP/SMET/C3 - Anemometer Readings Incorrect | twpsmet60sC3.b1 |
D060818.3 | TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa | twpsmet60sC3.b1 |
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/21/2002 | 0000 | 12/09/2002 | 0000 |
Subject: | TWP/MET/C3 - WIND DIRECTION SENSOR #2 MISALIGNED |
DataStreams: | twpsmet60sC3.a1, twpsmet60sC3.b1 |
Description: | Wind direction sensor #2 was misaligned and off by 15 degrees. |
Measurements: | twpsmet60sC3.a1:
twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
10/23/2002 | 0000 | 02/06/2003 | 2203 |
Subject: | TWP/SMET/C3 - Intermittent Pressure Sensor Failure |
DataStreams: | twpsmet60sC3.a1, twpsmet60sC3.b1 |
Description: | The Barometric Pressure sensor was intermittently failing at TWP site C3 Darwin. At times there were large pressure fluctuations occurring on the order of 24 mb in a few minutes. Times of large pressure fluctuations are readily apparent in the data. At these times the data should be carefully analyzed before determining it's usefulness. In between the periods of large fluctuations the readings appear to be stable. It is still suggested that independent pressure data be obtained for comparison before use. |
Measurements: | twpsmet60sC3.a1:
twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0628 | 06/03/2002 | 2359 |
Subject: | TWP/SMET/C3 - Data reported at 15sec rather than 60s intervals |
DataStreams: | twpsmet60sC3.a1, twpsmet60sC3.b1 |
Description: | Data in these data streams should be reports at 60sec intervals, but the logger program running during this time had the interval set to 15s. Users will have to average the 15sec data to get 60sec data. |
Measurements: | twpsmet60sC3.a1:
twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
01/01/2004 | 0000 | 05/17/2004 | 2259 |
Subject: | TWP/SMET/C3 - Rain blocked pressure port |
DataStreams: | twpsmet60sC3.b1 |
Description: | Atmospheric pressure values were suspect due to anomolous readings during heavy rain events. Suspect blockage of port due to rainfall. New logger and enclosure corrected problem. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/17/2004 | 2300 | 05/19/2004 | 0600 |
Subject: | TWP/SMET/C3 - Data missing due to Logger Change |
DataStreams: | twpsmet60sC3.b1 |
Description: | All Data from this site are missing during this time period due to a change of the collection system, logger and enclosure. Data resumed after change. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
03/12/2002 | 0628 | 05/17/2004 | 2300 |
Subject: | TWP/SMET/C3 - Negative rainfall rates |
DataStreams: | twpsmet60sC3.b1 |
Description: | Due to limitations of the Zeno dataloggers all voltage measurements from the optical rain gauge (ORG) were converted to rainfall rates. This resulted in negative values for rain rates. Any value below 0.10 mm/hr should be considered to be 0 mm/hr. New dataloggers alleviated this problem. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
02/09/2005 | 2310 | 02/09/2005 | 2320 |
Subject: | TWP/SMET/C3 - Data non-representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower mast was lowered for inspection. Data during this time is non-representative. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/14/2005 | 0230 | 04/19/2005 | 0050 |
Subject: | TWP/SMET/C3 - Instrument logger problem, Sensor failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | The wind_lower direction sensor failed on 4/14 and disagrees significantly with the wind_upper direction. The wind_lower sensor reports a constant value around 0-15 degrees since it failed. Testing showed that excitation channel two on the logger was the problem. Logger was replaced by spare. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
05/27/2004 | 0000 | 05/19/2005 | 2359 |
Subject: | TWP/SMET/C3 - Reprocess: Incorrect Units Listed for Vapor Pressure |
DataStreams: | twpsmet60sC3.b1 |
Description: | The netcdf file header incorrectly listed the units of Vapor Pressure as hPa when in fact they were kPa. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 0030 | 08/05/2005 | 0330 |
Subject: | TWP/SMET/C3 - Data not representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | Tower was lowered for radiometer change and characterization. Wind, temp and RH data not representative of local conditions. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
08/05/2005 | 0030 | 08/05/2005 | 0330 |
Subject: | TWP/SMET/C3 - Towrr lowered for sensor comparisons, data not representative |
DataStreams: | twpsmet60sC3.b1 |
Description: | The tower was lowered for radiometer change and comparisons. During the time period the data listed was not representative. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/15/2005 | 2215 |
Subject: | TWP/SMET/C3 - Upper wind sensor failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | A lightning strike destroyed the upper (#1) wind sensor. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/15/2005 | 0050 |
Subject: | TWP/SMET/C3 - RH Values Artificially High |
DataStreams: | twpsmet60sC3.b1 |
Description: | Lightning strike caused RH readings to be higher than reasonable (130%). |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
11/14/2005 | 0453 | 11/23/2005 | 0000 |
Subject: | TWP/SMET/C3 - Instrument failure |
DataStreams: | twpsmet60sC3.b1 |
Description: | A lightning strike caused a failure in the ORG. Rainfall was being continually recorded. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
04/27/2006 | 2330 | 05/02/2006 | 0201 |
Subject: | TWP/SMET/C3 - Reprocess: Windspeed2 calibration error |
DataStreams: | twpsmet60sC3.b1 |
Description: | The lower anemometer was replaced afer being removed in anticipation of cyclone Monica. The calibration coefficient was not changed during this period so the wind speeds are a little low (0.2%). |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
06/27/2006 | 0115 | 06/27/2006 | 0344 |
Subject: | TWP/SMET/C3 - Anemometer Readings Incorrect |
DataStreams: | twpsmet60sC3.b1 |
Description: | The Anemometer readings were incorrect due to the tower being lowered for maintenance. |
Measurements: | twpsmet60sC3.b1:
|
Start Date | Start Time | End Date | End Time |
---|---|---|---|
12/02/2003 | 0000 | 07/13/2006 | 2219 |
Subject: | TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa |
DataStreams: | twpsmet60sC3.b1 |
Description: | Barometric pressure data was converted from hPa to kPa in order to standardize the measurement units among ARM sites and to conform to accepted standard units determined by the scientific community. |
Measurements: | twpsmet60sC3.b1:
|