Data Quality Reports for Session: 101762 User: perez Completed: 09/27/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D000112.2TWP/SKYRAD - Manus PIR-G RGD1,2,3 bad 1999/10/25-2000/01/06twpskyradC1.00, twpskyrad20sC1.a1


DQRID : D000112.2
Start DateStart TimeEnd DateEnd Time
10/25/1999050001/06/20002359
Subject:
TWP/SKYRAD - Manus PIR-G RGD1,2,3 bad 1999/10/25-2000/01/06
DataStreams:twpskyradC1.00, twpskyrad20sC1.a1
Description:
Due to an error in the Manus SKYRAD data logger configuration, the
instantaneous values reported from the PIR-G (unshaded pyrgeometer) dome
thermistor are incorrect.  This only affects the instantaneous (10s, 30s
and 50s) data points.  The correct values were used in the calculation
of the PIR irradiance.  The affected values are obvious, as they are
round -26,000 ohms.  The bad data values are RGD1, RGD2 and RGD3, in
columns 11, 12 and 13 of the actual data (starting AFTER the "flags"
column).
Measurements:twpskyrad20sC1.a1:
  • Instantaneous PIR1 dome thermistor(pir1_dome_therm)

twpskyradC1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents



END OF DATA