Data Quality Reports for Session: 101770 User: perez Completed: 09/27/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040420.2SGP/SIRS/E5 - Dirt on PSP DomesgpsirsE5.00, sgpsirsE5.b1, sgpsirs20sE5.a0
D060630.13SGP/SIRS/E5 - Reprocess: Longwave Calibration errorsgpsirsE5.b1


DQRID : D040420.2
Start DateStart TimeEnd DateEnd Time
04/11/2004120005/05/20041645
Subject:
SGP/SIRS/E5 - Dirt on PSP Dome
DataStreams:sgpsirsE5.00, sgpsirsE5.b1, sgpsirs20sE5.a0
Description:
The PSP dome was dirty.  The magnitude of differences from derived 
             vs. measured values of downwelling shortwave were about 75 W/m^2. 
             Preventative maintenance on 4/21 cleaned instruments but they 
             became dirty again immediately afterward and continued to be
             so until the next PM on 5/5.
Measurements:sgpsirsE5.b1:
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Minima(down_short_hemisp_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer(short_direct_normal)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Standard
    Deviation(down_short_hemisp_std)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Maxima(down_short_hemisp_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer(down_short_hemisp)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Standard Deviation(short_direct_normal_std)

sgpsirs20sE5.a0:
  • Instantaneous Downwelling Hemispheric Shortwave, Unshaded Pyranometer Thermopile
    Voltage(inst_global)
  • Instantaneous Direct Normal Shortwave Irradiance, Pyrheliometer Thermopile
    Voltage(inst_direct_normal)

sgpsirsE5.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060630.13
Start DateStart TimeEnd DateEnd Time
03/10/2004203002/22/20061817
Subject:
SGP/SIRS/E5 - Reprocess: Longwave Calibration error
DataStreams:sgpsirsE5.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:sgpsirsE5.b1:
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)


Back To Table of Contents