Data Quality Reports for Session: 101785 User: perez Completed: 09/28/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D060630.17SGP/SIRS/E9 - Reprocess: Longwave Calibration errorsgpsirsE9.b1


DQRID : D060630.17
Start DateStart TimeEnd DateEnd Time
12/17/2003164002/07/20061730
Subject:
SGP/SIRS/E9 - Reprocess: Longwave Calibration error
DataStreams:sgpsirsE9.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:sgpsirsE9.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)


Back To Table of Contents



END OF DATA