Data Quality Reports for Session: 101804 User: maria Completed: 09/29/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030822.9TWP/MWR/C1 - min/max/delta values incorrecttwpmwrlosC1.b1
D040220.1TWP/MWR/C1 - wrong azimuthtwpmwrlosC1.a1, twpmwrlosC1.b1


DQRID : D030822.9
Start DateStart TimeEnd DateEnd Time
10/11/1996000002/09/20032359
Subject:
TWP/MWR/C1 - min/max/delta values incorrect
DataStreams:twpmwrlosC1.b1
Description:
The values of valid_min, valid_max, and valid_delta for fields tkxc and tknd were 
incorrect. They should be 303, 333, and 0.5 K, respectively
Measurements:twpmwrlosC1.b1:
  • Noise diode mount temperature(tknd)
  • Mixer kinetic (physical) temperature(tkxc)


Back To Table of Contents

DQRID : D040220.1
Start DateStart TimeEnd DateEnd Time
10/11/1996000002/18/20042250
Subject:
TWP/MWR/C1 - wrong azimuth
DataStreams:twpmwrlosC1.a1, twpmwrlosC1.b1
Description:
The MWR was initially installed at an azimuth angle defined as 180 degrees but the value 
in the configuration file was not changed from the default of 0 degrees. In examining 
photos taken during the installation of the AWS tower, I noticed that the MWR was rotated 
opposite the normal orientation. The value in the configuration file was changed to reflect 
the actual azimuth of the instrument.
Measurements:twpmwrlosC1.b1:
  • Actual Azimuth(actaz)

twpmwrlosC1.a1:
  • Actual Azimuth(actaz)


Back To Table of Contents



END OF DATA