Data Quality Reports for Session: 101821 User: perez Completed: 09/29/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040831.1SGP/SIRS/E1 - Reprocess: Incorrect shortwave radiometer calibration coefficientssgpsirsE1.00, sgpsirsE1.b1
D050205.1SGP/SIRS/E1 - broken cable on DSDH (aka DD) measurment instrument - Eppley 8-48.sgpsirsE1.00, sgpsirsE1.b1, sgpsirs20sE1.a0
D060309.1SGP/SIRS/E1 - Reprocess: incorrect calibration coefficients for US (upwelling shortwave)
instrument
sgpsirsE1.00, sgpsirsE1.b1
D060331.2SGP/SIRS/E1 - Intermittantly missing downwelling diffusesgpsirsE1.00, sgpsirsE1.b1, sgpsirs20sE1.a0
D060630.9SGP/SIRS/E1 - Reprocess: Longwave Calibration errorsgpsirsE1.b1


DQRID : D040831.1
Start DateStart TimeEnd DateEnd Time
02/10/2004205004/20/20042015
Subject:
SGP/SIRS/E1 - Reprocess: Incorrect shortwave radiometer calibration coefficients
DataStreams:sgpsirsE1.00, sgpsirsE1.b1
Description:
During the pyrgeometer instrument (PIR) changout a datalogger program was loaded with 
incorrect calibration coefficients for the shortwave instruments.  The affected measurements 
(instruments) were : downwelling diffuse (8-48), upwelling shortwave(PSP), direct normal 
(NIP) and downwelling shortwave (PSP)

Recalculate irradiances by determining original mV signal and reapplying the correct 
calibration coefficients.  For each one minute irradiance value perform the following 
calculations by instrument:
DD(corrected) = (DD/107.24)*109.01
US(corrected) = (US/113.19)*111.47
NIP(corrected)= (NIP/120.17)*114.82
DS(corrected) = (DS/118.46)*116.51
Measurements:sgpsirsE1.b1:
  • Shortwave Direct Normal Irradiance, Pyrheliometer(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Maxima(down_short_hemisp_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Standard
    Deviation(down_short_hemisp_std)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Minima(down_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(up_short_hemisp_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Standard Deviation(short_direct_normal_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer(up_short_hemisp)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(up_short_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer(down_short_hemisp)

sgpsirsE1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D050205.1
Start DateStart TimeEnd DateEnd Time
12/30/2004210002/08/20052055
Subject:
SGP/SIRS/E1 - broken cable on DSDH (aka DD) measurment instrument - Eppley 8-48.
DataStreams:sgpsirsE1.00, sgpsirsE1.b1, sgpsirs20sE1.a0
Description:
Occasionally, the downwelling shortwave diffuse hemispheric (DSDH) readings spike to 
unreasonably low levels, causing derived downwelling shortwave hemispheric (DSH) readings to 
spike well below the actual DSH readings.  Technicians repaired the DSDH instument (an 
Eppley 8-48) cable on 02/08/2005 2055GMT.  There have been no dropouts since that time.
Measurements:sgpsirs20sE1.a0:
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)

sgpsirsE1.b1:
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)

sgpsirsE1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060309.1
Start DateStart TimeEnd DateEnd Time
02/21/2006200803/07/20061950
Subject:
SGP/SIRS/E1 - Reprocess: incorrect calibration coefficients for US (upwelling shortwave) 
instrument
DataStreams:sgpsirsE1.00, sgpsirsE1.b1
Description:
During the changeover from 4 to 1 coefficient for the PIRs at SIRS/E1 (PIF) the datalogger 
program was programmed with the wrong calibration coefficient and serial number for the 
US PSP instrument measuring the Upwelling Hemispheric Shortwave radiation. The 
calibration coefficient was 114.43 W/m2/mV (instrument 30618F3) when it should have been 
104.47W/m2/mV (instrument 29618F3)
Measurements:sgpsirsE1.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(up_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer(up_short_hemisp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(up_short_hemisp_max)

sgpsirsE1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060331.2
Start DateStart TimeEnd DateEnd Time
10/26/2005163004/04/20062005
Subject:
SGP/SIRS/E1 - Intermittantly missing downwelling diffuse
DataStreams:sgpsirsE1.00, sgpsirsE1.b1, sgpsirs20sE1.a0
Description:
Starting on or about 10/26/2005 the downwelling diffuse hemispheric irradiance (Eppley 
8-48) instrument had intermittant missing values. Technicians found a loose wire on the 
instrument connector and repaired it on 4/4/2006. 

Use global shortwave hemispheric (PSP) component - normal incident component (CosZA*NIP) 
to calulate diffuse component.  Comparison of absolute difference between actual and 
computed diffuse component values and diffuse standard deviations > 1-2 W/m2 can help filter 
the noisy diffuse data.  There is no standard method.
Measurements:sgpsirs20sE1.a0:
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)

sgpsirsE1.b1:
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)

sgpsirsE1.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D060630.9
Start DateStart TimeEnd DateEnd Time
02/10/2004205502/21/20062007
Subject:
SGP/SIRS/E1 - Reprocess: Longwave Calibration error
DataStreams:sgpsirsE1.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                The data collected while the incorrect procedures were in place are 
being reprocessing to remove the calibration bias.  Note, this also affected standard 
deviation, maximum and minimum data fields.
Measurements:sgpsirsE1.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)


Back To Table of Contents



END OF DATA