Data Quality Reports for Session: 102180 User: perez Completed: 10/18/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D040130.1SGP/AOS/C1 - Ozone Instrument removed for repairsgpaosC1.a0, sgpaosauxC1.a0
D041026.3SGP/AOS/C1 - PCASP instrument problemsgpaosC1.a0, sgpaosauxC1.a0
D041026.4SGP/AOS/C1 - Ozone instrument removed for repairsgpaosC1.a0, sgpaosauxC1.a0
D041026.5SGP/AOS/C1 - Condensation Particle Counter failuresgpaosC1.a0, sgpaosauxC1.a0
D041026.6SGP/AOS/C1 - Wet nephelometer problemssgpaosC1.a0, sgpaosauxC1.a0
D050120.2SGP/AOS/C1 - Failed RH sensorsgpaosC1.a0
D980528.1SGP/AOS/C1 - Optical Particle Counter (PCASP-X) downsgpaosC1.a0, sgpaosauxC1.a0


DQRID : D040130.1
Start DateStart TimeEnd DateEnd Time
11/21/2003163601/27/20042208
Subject:
SGP/AOS/C1 - Ozone Instrument removed for repair
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The ozone instrument was removed for calibration on 2003 11/21.  It was shipped to Boulder 
and was cleaned and calibrated.  It was sent back to SGP on 2003 12/02, but was damaged 
in transit by FedEx.  The instrument was reinstalled on 2003 12/03 but was found to be 
defective.  Large spikes were observed in the data.  Afetr a few days the instrument was 
removed and shipped to a vendor for repair.  The ozone monitor was returned to service 
after repair at 2004 01/27 (2208 GMT).
Measurements:sgpaosauxC1.a0:
  • Ozone sample frequency(O3_Samp)
  • Ozone offset value(O3_Offset)
  • Ozone sample pressure(O3_Press)
  • Ozone span value(O3_Span)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Ozone sample temperature(O3_Temp)
  • Ozone line temperature set point(O3_SetPt)
  • Ozone span concentration(O3_Spn)
  • Ozone zero concentration(O3_Zro)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • unknown(O3_Cont)
  • Ozone mode(O3_Mode)

sgpaosC1.a0:
  • Ozone concentration(Ozone)


Back To Table of Contents

DQRID : D041026.3
Start DateStart TimeEnd DateEnd Time
01/10/2004000010/26/20041200
Subject:
SGP/AOS/C1 - PCASP instrument problem
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
PCASP (optical particle counter) gave unusually low aerosol signal counts despite numerous 
attempts to fix the instrument by the mentor and manufacturer.
Measurements:sgpaosauxC1.a0:
  • OPC sheath(OPCSheath)
  • OPC sample flow(OPCFlow)
  • Analog signal from PMS PCASP(OPCLaserVolt)

sgpaosC1.a0:
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)


Back To Table of Contents

DQRID : D041026.4
Start DateStart TimeEnd DateEnd Time
11/21/2003160001/28/20040000
Subject:
SGP/AOS/C1 - Ozone instrument removed for repair
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The ozone monitor was removed and sent back to the manufacturer for repair.
Measurements:sgpaosauxC1.a0:
  • Ozone sample frequency(O3_Samp)
  • Ozone offset value(O3_Offset)
  • Ozone sample pressure(O3_Press)
  • Ozone span value(O3_Span)
  • Ozone line temp. from thermocouple attached to heater(O3_LinHeater)
  • Ozone sample temperature(O3_Temp)
  • Ozone line temperature set point(O3_SetPt)
  • Ozone span concentration(O3_Spn)
  • Ozone zero concentration(O3_Zro)
  • Ozone line sample temperature (omega temp. sensor)(O3_LinTemp)
  • unknown(O3_Cont)
  • Ozone mode(O3_Mode)

sgpaosC1.a0:
  • Ozone concentration(Ozone)


Back To Table of Contents

DQRID : D041026.5
Start DateStart TimeEnd DateEnd Time
07/28/2004140008/27/20041540
Subject:
SGP/AOS/C1 - Condensation Particle Counter failure
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The CPC signal declined to a low value. The instrument was shipped back to the mentor, 
cleaned and repaired. The instrument inlet was clogged.
Measurements:sgpaosauxC1.a0:
  • TSI CPC saturator temperature(CPCSaturT)
  • TSI CPC condensor and saturator temperature difference(CPCDeltT)
  • TSI CPC condensor temperature(CPCCondT)

sgpaosC1.a0:
  • CPC Particle concentration(CPCPartConc)


Back To Table of Contents

DQRID : D041026.6
Start DateStart TimeEnd DateEnd Time
01/24/2004000002/18/20041800
Subject:
SGP/AOS/C1 - Wet nephelometer problems
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The humidifier tube was coated with mold and had to be replaced. This affects the wet 
nepheleometer measurements and the calculations of the aerosol hygroscopic growth.

The humidifer was cleaned and replaced. A filter was added to the water circulation to 
minimize mold.
Measurements:sgpaosauxC1.a0:
  • TSI High RH Neph. 550 nm back scattering coefficient for zero gas(GrnBScatZero_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for zero gas(GrnTScatZero_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for cal. gas(RedTScatSpn_HRH)
  • TSI High RH Neph. lamp voltage(TSINephLV_HRH)
  • TSI High RH Neph. 550 nm total scattering coefficient for cal. gas(GrnTScatSpn_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for cal. gas(BluTScatSpn_HRH)
  • TSI High RH Neph. 700 nm total scattering coefficient for zero gas(RedTScatZero_HRH)
  • TSI High RH Neph. 550 nm rayleigh scattering coefficient for zero gas(GrnRScatZero_HRH)
  • TSI High RH Neph. lamp current(TSINephLC_HRH)
  • TSI High RH Neph. 550 nm back scattering coefficient for cal. gas(GrnBScatSpn_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for zero gas(BluBScatZero_HRH)
  • TSI High RH Neph. 450 nm total scattering coefficient for zero gas(BluTScatZero_HRH)
  • TSI High RH Neph. 550nm reference intensity(TSISensGrn_HRH)
  • TSI High RH Neph. 450 nm rayleigh scattering coefficient for zero gas(BluRScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for zero gas(RedBScatZero_HRH)
  • TSI High RH Neph. 700 nm back scattering coefficient for cal. gas(RedBScatSpn_HRH)
  • TSI High RH Neph. 700 nm rayleigh scattering coefficient for zero gas(RedRScatZero_HRH)
  • TSI High RH Neph. 450 nm back scattering coefficient for cal. gas(BluBScatSpn_HRH)

sgpaosC1.a0:
  • TSI High RH Neph. 700 nm total scat. coef. at 1 um(RedTScatCoef_1um_HRH)
  • Temp. downstream of humidograph(HGdownstream_T)
  • TSI High RH Neph. Pressure(TSINephPres_HRH)
  • TSI High RH Neph. Sample Temperature(TSINephTSamp_HRH)
  • TSI High RH Neph. 700 nm backscat. coef. at 10 um(RedBScatCoef_10um_HRH)
  • RH upstream of humidograph(HGupstream_RH)
  • TSI High RH Neph. 700 nm backscat. coef. at 1 um(RedBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm backscat. coef. at 1 um(GrnBScatCoef_1um_HRH)
  • Temp. upstream of humidograph(HGupstream_T)
  • TSI High RH Neph. 550 nm backscat. coef. at 10 um(GrnBScatCoef_10um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 1 um(GrnTScatCoef_1um_HRH)
  • TSI High RH Neph. inlet temperature(TSINephTin_HRH)
  • TSI High RH Neph. 700 nm total scat. coef. at 10 um(RedTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm backscat. coef. at 1 um(BluBScatCoef_1um_HRH)
  • TSI High RH Neph. 550 nm total scat. coef. at 10 um(GrnTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 10 um(BluTScatCoef_10um_HRH)
  • TSI High RH Neph. 450 nm total scat. coef. at 1 um(BluTScatCoef_1um_HRH)
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)
  • RH downstream of humidograph(HGdownstream_RH)
  • TSI High RH Neph. 450 nm backscat. coef. at 10 um(BluBScatCoef_10um_HRH)


Back To Table of Contents

DQRID : D050120.2
Start DateStart TimeEnd DateEnd Time
01/01/2004000005/01/20052359
Subject:
SGP/AOS/C1 - Failed RH sensor
DataStreams:sgpaosC1.a0
Description:
The RH sensor inside the humidified nephelometer began to slowly decline in value in the 
upper RH range beginning in early 2004 and becoming more noticable poor on 20041117. The 
sensor currently reaches a high RH value of about 60% when the actual RH is closer to 80%.

The sensor was replaced in the Spring of 2005, but the problem remained.  Effective 
1/1/2004, this sensor is not used to calculate scientific data and is therefore essentially 
abandoned in the field.  An end-date of 20050501 is being used as a placeholder.
Measurements:sgpaosC1.a0:
  • TSI High RH Neph. relative humidity(TSINephRHSamp_HRH)


Back To Table of Contents

DQRID : D980528.1
Start DateStart TimeEnd DateEnd Time
07/01/1997093607/03/19971556
08/01/1997132908/08/19971348
05/05/1998092606/16/19981505
06/09/1999082407/26/19992205
09/08/2000042610/13/20001630
05/17/2002060205/20/20051847
05/23/2002201105/31/20021258
09/09/2002000010/07/20021905
Subject:
SGP/AOS/C1 - Optical Particle Counter (PCASP-X) down
DataStreams:sgpaosC1.a0, sgpaosauxC1.a0
Description:
The AOS Optical Particle Spectrometer (PCASP-X sizing probe) was down during these 
periods.  This lowers the laser reference voltage and results in inaccurate sizing.
Measurements:sgpaosauxC1.a0:
  • OPC sheath(OPCSheath)
  • OPC sample flow(OPCFlow)
  • Analog signal from PMS PCASP(OPCLaserVolt)

sgpaosC1.a0:
  • PMS PCASP Chan. 7 (0.23 micrometers < Dp < 0.26)(pa23_p26conc)
  • PMS PCASP Chan. 26 (3.00 micrometers < Dp < 3.50)(pa300_p350conc)
  • PMS PCASP Chan. 28 (4.00 micrometers < Dp < 5.00)(pa400_p500conc)
  • PMS PCASP Chan. 15 (0.70 micrometers < Dp < 0.80)(pa70_p80conc)
  • PMS PCASP Chan. 25 (2.60 micrometers < Dp < 3.00)(pa260_p300conc)
  • PMS PCASP Chan. 0 (Dp > 10 micrometers)(Pa1000Conc)
  • PMS PCASP Chan. 2 (0.12 micrometers < Dp < 0.14)(pa12_p14conc)
  • PMS PCASP Chan. 30 (6.50 micrometers < Dp < 8.00)(pa650_p800conc)
  • PMS PCASP Chan. 6 (0.20 micrometers < Dp < 0.23)(pa20_p23conc)
  • PMS PCASP Chan. 19 (1.30 micrometers < Dp < 1.40)(pa130_p140conc)
  • PMS PCASP Chan. 17 (0.90 micrometers < Dp < 1.00)(pa90_p100conc)
  • PMS PCASP Chan. 18 (1.00 micrometers < Dp < 1.30)(pa100_p130conc)
  • PMS PCASP Chan. 3 (0.14 micrometers < Dp < 0.16)(pa14_p16conc)
  • PMS PCASP Chan. 31 (8.00 micrometers < Dp < 10.0)(pa800_p1000conc)
  • PMS PCASP Chan. 1 1/cm^3 (0.10 micrometers < Dp < 0.12)(pa10_p12conc)
  • PMS PCASP Chan. 23 1/cm^3 (2.00 micrometers < Dp < 2.30)(pa200_p230conc)
  • PMS PCASP Chan. 24 (2.30 micrometers < Dp < 2.60)(pa230_p260conc)
  • PMS PCASP Chan. 21 (1.60 micrometers < Dp < 1.80)(pa160_p180conc)
  • PMS PCASP Chan. 22 (1.80 micrometers < Dp < 2.00)(pa180_p200conc)
  • PMS PCASP Chan. 9 (0.30 micrometers < Dp < 0.35)(pa30_p35conc)
  • PMS PCASP Chan. 27 (3.50 micrometers < Dp < 4.00)(pa350_p400conc)
  • PMS PCASP Chan. 13 (0.50 micrometers < Dp < 0.60)(pa50_p60conc)
  • PMS PCASP Chan. 4 (0.16 micrometers < Dp < 0.18)(pa16_p18conc)
  • PMS PCASP Chan. 20 (1.40 micrometers < Dp < 1.60)(pa140_p160conc)
  • PMS PCASP Chan. 8 (0.26 micrometers < Dp < 0.30)(pa26_p30conc)
  • PMS PCASP Chan. 5 (0.18 micrometers < Dp < 0.20)(pa18_p20conc)
  • PMS PCASP Chan. 14 (0.60 micrometers < Dp < 0.70)(pa60_p70conc)
  • PMS PCASP Chan. 10 (0.35 micrometers < Dp < 0.40)(pa35_p40conc)
  • PMS PCASP Chan. 11 (0.35 micrometers < Dp < 0.40)(pa40_p45conc)
  • PMS PCASP Chan. 16 (0.80 micrometers < Dp < 0.90)(pa80_p90conc)
  • PMS PCASP Chan. 12 (0.45 micrometers < Dp < 0.50)(pa45_p50conc)
  • PMS PCASP Chan. 29 (5.00 micrometers < Dp < 6.50)(pa500_p650conc)


Back To Table of Contents