Data Quality Reports for Session: 102341 User: dmechem Completed: 10/27/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D050614.1SGP/VCEIL/C1 - poor signalsgpvceil25kC1.b1


DQRID : D050614.1
Start DateStart TimeEnd DateEnd Time
05/05/2005184204/19/20061713
Subject:
SGP/VCEIL/C1 - poor signal
DataStreams:sgpvceil25kC1.b1
Description:
When parameters were changed (set noise_h2_compensation to "ON" and units to "METERS") to 
correct a low amplitude problem, the backscatter plots began displaying horizontal 
stripes indicating detector "ringing". However, the cloud base heights are not affected. The 
problem was corrected when the instrument was replaced.
Measurements:sgpvceil25kC1.b1:
  • Detection status. See details(detection_status)
  • Aerosol volume backscattering coefficient at 355 nm(backscatter)


Back To Table of Contents



END OF DATA