Data Quality Reports for Session: 102385 User: mstreibel Completed: 10/31/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D051115.1TWP/SMET/C3 - Upper wind sensor failuretwpsmet60sC3.b1
D051117.2TWP/SMET/C3 - RH Values Artificially Hightwpsmet60sC3.b1
D051117.3TWP/SMET/C3 - Instrument failuretwpsmet60sC3.b1
D060426.8TWP/SMET/C3 - ORG data noisy due to degraded groundtwpsmet60sC3.b1
D060818.3TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPatwpsmet60sC3.b1
D061011.4TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge addedtwpsmet60sC3.b1


DQRID : D051115.1
Start DateStart TimeEnd DateEnd Time
11/14/2005045311/15/20052215
Subject:
TWP/SMET/C3 - Upper wind sensor failure
DataStreams:twpsmet60sC3.b1
Description:
A lightning strike destroyed the upper (#1) wind sensor.
Measurements:twpsmet60sC3.b1:
  • Wind #1 direction vector average(wind1_dir_vec_avg)
  • Wind #1 speed minimum(wind1_spd_min)
  • Wind #1 speed maximum(wind1_spd_max)
  • Wind #1 speed standard deviation(wind1_spd_sd)
  • Wind #1 speed maximum gust(wind1_spd_max_gust)
  • Wind #1 speed arithmetic average(wind1_spd_arith_avg)
  • Wind #1 speed vector average(wind1_spd_vec_avg)
  • Wind #1 direction standard deviation(wind1_dir_sd)


Back To Table of Contents

DQRID : D051117.2
Start DateStart TimeEnd DateEnd Time
11/14/2005045311/15/20050050
Subject:
TWP/SMET/C3 - RH Values Artificially High
DataStreams:twpsmet60sC3.b1
Description:
Lightning strike caused RH readings to be higher than reasonable (130%).
Measurements:twpsmet60sC3.b1:
  • Vapor pressure standard deviation(vappress_sd)
  • Relative humidity mean(relh_mean)
  • Relative humidity maximum(relh_max)
  • Vapor pressure mean(vappress_mean)
  • Std Dev of Relative Humidity or Hardware Error Code(relh_sd)
  • Relative humidity minimum(relh_min)


Back To Table of Contents

DQRID : D051117.3
Start DateStart TimeEnd DateEnd Time
11/14/2005045311/23/20050000
Subject:
TWP/SMET/C3 - Instrument failure
DataStreams:twpsmet60sC3.b1
Description:
A lightning strike caused a failure in the ORG.  Rainfall was being continually recorded.
Measurements:twpsmet60sC3.b1:
  • Precipitation mean(precip_mean)
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D060426.8
Start DateStart TimeEnd DateEnd Time
12/01/2005000008/24/20062359
Subject:
TWP/SMET/C3 - ORG data noisy due to degraded ground
DataStreams:twpsmet60sC3.b1
Description:
SMET Rain Data began to be noisy during December 2005 and the problem slowly became worse 
over time so that in February 2006 the ORG was reporting with a constant background value 
of around 0.10 mm/hr rain rate.  ORG voltage showed a near constant value above the 85mV 
threshold.  The ground for the ORG had become degraded and was lost due to corrosion.
Measurements:twpsmet60sC3.b1:
  • Precipitation mean(precip_mean)
  • Precipitation maximum(precip_max)
  • Precipitation standard deviation(precip_sd)
  • precipitation minimum(precip_min)


Back To Table of Contents

DQRID : D060818.3
Start DateStart TimeEnd DateEnd Time
12/02/2003000007/13/20062219
Subject:
TWP/SMET/C3 - Reprocess: Barometric Data Changed from hPa to kPa
DataStreams:twpsmet60sC3.b1
Description:
Barometric pressure data was converted from hPa to kPa in order to standardize the 
measurement units among ARM sites and to conform to accepted standard units determined by the 
scientific community.
Measurements:twpsmet60sC3.b1:
  • Atmospheric pressure(atmos_pressure)


Back To Table of Contents

DQRID : D061011.4
Start DateStart TimeEnd DateEnd Time
03/12/2002000009/26/20061435
Subject:
TWP/SMET/C3 - Reprocess: Tipping bucket rain gauge added
DataStreams:twpsmet60sC3.b1
Description:
A tipping bucket rain gauge was added to the SMET suite of instruments effective 20060926. 
 Two new variables (count and rain amount) were added at the end of the twpsmet60sC1.b1 
datastream.  When reprocessing of this datastream occurs these variables will be added 
with -9999 (missing) as the value for dates prior to gauge installation.
Measurements:twpsmet60sC3.b1:
  • Time offset from base_time(base_time)


Back To Table of Contents



END OF DATA