Data Quality Reports for Session: 102427 User: yan Completed: 11/01/2006


TABLE OF CONTENTS

DQR IDSubjectData Streams Affected
D030110.36SGP/SIRS/E10 - Noisy UIRsgpsirsE10.a2, sgpsirsE10.b1, sgpsirs60sE10.a1
D030701.1SGP/SIRS/E10 - Noisy US(upwelling shortwave) data stream problemsgpsirsE10.00, sgpsirsE10.b1, sgpsirs20sE10.a0
D031027.1SGP/SIRS/E10 - Noisy UIR (upwelling longwave)data stream problemsgpsirsE10.00, sgpsirsE10.b1, sgpsirs20sE10.a0
D040831.4SGP/SIRS/E10 - Reprocess: Incorrect shortwave radiometer calibration coefficientssgpsirsE10.00, sgpsirsE10.b1
D041206.1SGP/SIRS/E10 - tracker failuresgpsirsE10.00, sgpsirsE10.b1, sgpsirs20sE10.a0
D060630.18SGP/SIRS/E10 - Reprocessed: Longwave Calibration errorsgpsirsE10.b1


DQRID : D030110.36
Start DateStart TimeEnd DateEnd Time
01/01/2002000011/08/20021900
Subject:
SGP/SIRS/E10 - Noisy UIR
DataStreams:sgpsirsE10.a2, sgpsirsE10.b1, sgpsirs60sE10.a1
Description:
Upwelling Infrared irradiance signals noisy resulting from problems with pyrgeometer 
signal cable connections.
Measurements:sgpsirsE10.b1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Maxima(up_long_hemisp_max)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Standard Deviation(up_long_hemisp_std)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Minima(up_long_hemisp_min)

sgpsirsE10.a2:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Standard Deviation(up_long_hemisp_std)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Maxima(up_long_hemisp_max)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Minima(up_long_hemisp_min)

sgpsirs60sE10.a1:
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Maxima(up_long_hemisp_max)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Minima(up_long_hemisp_min)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Standard Deviation(up_long_hemisp_std)


Back To Table of Contents

DQRID : D030701.1
Start DateStart TimeEnd DateEnd Time
06/22/2003014607/02/20031400
Subject:
SGP/SIRS/E10 - Noisy US(upwelling shortwave) data stream problem
DataStreams:sgpsirsE10.00, sgpsirsE10.b1, sgpsirs20sE10.a0
Description:
Beginning 6/22, upwelling shortwave values frequently dropped well 
                below 0 W/m^2 (occasionally dropping below -50 W/m^2) during the
                nighttime hours.  Daytime values were excessively high and noisy.
                Instrument wiring was suspected and was corrected 07/02/2003.
Measurements:sgpsirsE10.b1:
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer(up_short_hemisp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(up_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(up_short_hemisp_min)

sgpsirsE10.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)

sgpsirs20sE10.a0:
  • Instantaneous Upwelling Shortwave Irradiance, Inverted Pyranometer Thermopile
    Voltage(inst_up_short_hemisp)


Back To Table of Contents

DQRID : D031027.1
Start DateStart TimeEnd DateEnd Time
06/29/2003150008/06/20030500
Subject:
SGP/SIRS/E10 - Noisy UIR (upwelling longwave)data stream problem
DataStreams:sgpsirsE10.00, sgpsirsE10.b1, sgpsirs20sE10.a0
Description:
Beginning 6/29, upwelling longwave (UIR) values frequently
                 exceeded quality control boundaries and appeared excessively 
                 noisy. Instrument wiring was suspected.  Data remained noisy 
                 until 08/06/2003 0500GMT when the UIR data returned to normal.
Measurements:sgpsirsE10.b1:
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Temperature, Pyrgeometer(inst_up_long_dome_temp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Maxima(up_long_hemisp_max)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Standard Deviation(up_long_hemisp_std)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer,
    Minima(up_long_hemisp_min)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Temperature, Pyrgeometer(inst_up_long_case_temp)

sgpsirs20sE10.a0:
  • Instantaneous Upwelling Pyrgeometer Dome Thermistor Resistance, Pyrgeometer(inst_up_long_dome_resist)
  • Instantaneous Upwelling Pyrgeometer Thermopile Voltage, Pyrgeometer(inst_up_long_hemisp_tp)
  • Instantaneous Upwelling Pyrgeometer Case Thermistor Resistance, Pyrgeometer(inst_up_long_case_resist)

sgpsirsE10.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D040831.4
Start DateStart TimeEnd DateEnd Time
02/20/2004163004/21/20041425
Subject:
SGP/SIRS/E10 - Reprocess: Incorrect shortwave radiometer calibration coefficients
DataStreams:sgpsirsE10.00, sgpsirsE10.b1
Description:
During the pyrgeometer instrument (PIR) changeout a datalogger program was loaded with 
incorrect calibration coefficients for the shortwave instruments.  The affected measurements 
(instruments) were: downwelling diffuse (8-48), upwelling shortwave (PSP), direct normal 
(NIP) and downwelling shortwave (PSP)

Recalculate irradiances by determining original mV signal and reapplying the correct 
calibration coefficients.  For SGP/SIRS/E10 and each one minute irradiance value perform the 
following calculations by instrument:
DD(corrected) = (DD/107.24)*114.63
US(corrected) = (US/113.19)*117.46
NIP(corrected)= (NIP/120.17)*119.26
DS(corrected) = (DS/118.46)*122.78
Measurements:sgpsirsE10.b1:
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(up_short_hemisp_std)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(up_short_hemisp_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Maxima(down_short_hemisp_max)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer(up_short_hemisp)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer(down_short_hemisp)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Standard
    Deviation(down_short_hemisp_std)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Standard Deviation(short_direct_normal_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer(short_direct_normal)
  • Downwelling Shortwave Hemispheric Irradiance, Ventilated Pyrgeometer, Minima(down_short_hemisp_min)
  • Upwelling (10 meter) Shortwave Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(up_short_hemisp_min)

sgpsirsE10.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)


Back To Table of Contents

DQRID : D041206.1
Start DateStart TimeEnd DateEnd Time
11/04/2004180001/26/20051600
Subject:
SGP/SIRS/E10 - tracker failure
DataStreams:sgpsirsE10.00, sgpsirsE10.b1, sgpsirs20sE10.a0
Description:
During the night of 11/05-06 2004 the tracker failed at this location. During CM on 12/01, 
power cycling returned the tracker to service. However, tracker was not properly 
tracking in the AM. Tracker failed again sometime near 01/01/2005.  CM on reloaded the tracker 
program and realigned the tracker on 01/26/2005 1600GMT.  It has been working correctly 
since this time.
Measurements:sgpsirsE10.b1:
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Minima(short_direct_normal_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Maxima(short_direct_normal_max)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Standard Deviation(down_short_diffuse_hemisp_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer(down_short_diffuse_hemisp)
  • Shortwave Direct Normal Irradiance, Pyrheliometer, Standard Deviation(short_direct_normal_std)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Minima(down_short_diffuse_hemisp_min)
  • Shortwave Direct Normal Irradiance, Pyrheliometer(short_direct_normal)
  • Downwelling Shortwave Diffuse Hemispheric Irradiance, Ventilated Pyranometer,
    Maxima(down_short_diffuse_hemisp_max)

sgpsirsE10.00:
  • Raw data stream - documentation not supported(Raw data stream - documentation not supported)

sgpsirs20sE10.a0:
  • Instantaneous Uncorrected Downwelling Shortwave Diffuse, Shaded Pyranometer
    Thermopile Voltage(inst_diffuse)
  • Instantaneous Direct Normal Shortwave Irradiance, Pyrheliometer Thermopile
    Voltage(inst_direct_normal)


Back To Table of Contents

DQRID : D060630.18
Start DateStart TimeEnd DateEnd Time
02/20/2004173002/07/20062130
Subject:
SGP/SIRS/E10 - Reprocessed: Longwave Calibration error
DataStreams:sgpsirsE10.b1
Description:
Modified pyrgeometer calibration procedures were implemented beginning in December 2003.  
These modified procedures introduced a calibration bias in the longwave data.  The 
previous procedures were re-implemented at all sites between December 2005 and February 2006 to 
restore proper calibrations.                                                             
                
The data collected while the incorrect procedures were in place have been reprocessed to 
remove the calibration bias.  The reprocessed 60 second averaged data are based on 3 
instantaneous 20 second data records rather than on 60 1 second instantaneous data records. 
Still, these data are considered far superior to the originally processed data.  The 
reprocessed data were archived in August 2006.
Measurements:sgpsirsE10.b1:
  • Upwelling Longwave Hemispheric Net Infrared(up_long_netir)
  • Upwelling (10 meter) Longwave Hemispheric Irradiance, Ventilated Pyrgeometer(up_long_hemisp)
  • Downwelling Longwave Hemispheric Irradiance, Shaded Pyrgeometer(down_long_hemisp_shaded)
  • Downwelling Longwave Hemispheric Net Infrared(down_long_netir)


Back To Table of Contents



END OF DATA